Wiring for RS–232
Module
• If the module has an RS–232 output, see Figure 2-3.
• Table 2-2 lists pin designations for the 9-pin subminiature D connector.
Figure 2-3
Wiring Terminals for Micro-Ion E Module with RS–232 Output
Grounding
Table 2-2
Wiring Pin Designations for RS–232 Module
Pin designation
Function
Pin number
Chassis ground
Ground connection for module chassis
1
RS–232 TDX
• Pins 2, 3, and 5 enable RS–232 digital communication
• Pins, 2, 3, and 5 connect to computer communication port
2
RS–232 RDX
3
RS–232 common
5
Not used
Reserved
4
Not used
Reserved
6
Not used
Reserved
7
24 Vdc power supply input (+)
Positive connection for 24 Vdc power supply
8
24 Vdc power supply ground (–)
Negative connection for 24 Vdc power supply
9
RS–232 common
Not used
RS–232 RDX
RS–232 TDX
Chassis ground
24 Vdcpower ground (–)
24 Vdcpower input (+)
Not used
Not used
Male pins
Improper grounding could cause severe product failure
or personal injury.
Follow ground network requirements for the facility.
•
Maintain all exposed conductors at earth ground.
•
Ground the module housing to the vacuum chamber as
illustrated below.
•
Make sure the vacuum port to which the module is mounted
is properly grounded.
Содержание Micro-Ion 354 E Series
Страница 2: ......
Страница 10: ...10 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 24: ...24 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 42: ...Chapter 6 42 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 46: ...46 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 52: ...52 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 53: ......