CI606-Q Thermal Barrier Performance
Dimensions: (230 mm/230 mm/100 mm)
Table of Results:
T initial =20°C
T initial=30°C
T oven (°C)
T oven (°C)
Time to reach 60°C
Time to reach 60°C
100
150
200
250
300
100
150
200
250
300
180 minutes
95 minutes
68 minutes
58 minutes
52 minutes
130 minutes
77 minutes
60 minutes
51 minutes
46 minutes
Note: T initial is the initial temperature of the thermal barrier before placing it in the Oven.