Recirculating Chiller
Part No 09931/Issue
10
Page
14
April
7
Maintenance
All Grant laboratory products are designed to comply with IEC1010-1 and can be flash tested.
Some are fitted with radio frequency interference suppressers. Therefore it is recommended that
only a d.c. test is performed.
Routine maintenance:
Check monthly that dust is removed from the grilles.
Check monthly that the line strainer which is fitted on the inlet pipe is clean, particularly when
foreign particles can enter the pumped liquid. Switch off, and shut off and drain any head of
liquid. Undo the strainer retaining nut on the underside of the brass fitting, remove the strainer
and clean it. Refit by reversing the procedure.
Inspect and replace where necessary any worn or damaged hoses.
7.1
Cleaning
The case can be cleaned with a damp cloth after disconnection. Do not use solvents.
Before using any decontamination or cleaning method except that recommended, check with our
service department, or in other countries with our distributor, that the proposed method will not
damage the equipment
7.2
Replacement of fuses
Fuse replacement should only be carried out by a competent person.
7.2.1 RC350G and RC400G
Disconnect the unit from the power supply.
Remove the IEC plug from the socket in the back of the bath.
Press down the fuse drawer catch.
Pull out the fuse drawer, check and replace with the correct fuses if necessary.
The fuses should be 1.25 x 0.25 inch ceramic quick acting, rated: 10AF 240V.
Push back the drawer and replace the IEC plug.
7.2.2 RC1400G and RC3000G
Disconnect the unit from the power supply.
Locate the fuses on the control unit rear panel to the power cable inlet.
Use a screwdriver of correct size to turn the fuse holder cap a quarter turn anti-clockwise to
release the fuse holder cap.
Check and replace fuse if necessary.
Replace with correct fuses.
The fuse should be 1.25 x 0.25 inch ceramic quick acting rated 15AF 240V.
Replace the fuse holder cap.
8
Guarantee
When used under laboratory conditions and according to these instructions this chiller is
guaranteed for three years against faulty materials or workmanship.
9 Service
For service, ret
urn for repair to our Service Department or, in other countries, to our distributor
.
Grant Instruments (Cambridge) Ltd
Shepreth
Cambridgeshire.
SG8 6GB
England
Tel: +44 01763 260811
Fax: +44 01763 262410