Make sure that the weighing pan is empty. When the AD value is stable, press
the key.
3. Configuration of the value of the calibration weight.
Display
Descriptions
XXXXXX
XXXXX
Pnt.CAL
The first row shows the AD value, the second row
shoes the calibration weight (Unit: grams or lb/1000)
and the third row shows “Pnt.CAL”.
The default value is 1/3 of the scale’s capacity. Use
the numerical keypad to enter the desired value.
Place the calibration weight on the weighing pan. Use the numerical keypad to
enter its value in grams.
When the AD value is stable, press the key to conclude the calibration
process.
LINEARITY CALIBRATION (CAL 5)
Note: This section can only be carried out by professionals.
Remove the JP3 jumper before proceeding to calibration.
In the Calibration menu, press the C key to navigate between the different
options and press the key to confirm.
1. Configure maximum capacity
Display
Descriptions
Kg
Lb
Viewed max. capacity varies with the default wt unit.
LoAd
03
LoAd
06
Configure maximum capacity as 3 kg/6lb.
LoAd
06
LoAd
12
Configure maximum capacity as 6 kg/12lb.
LoAd
15
LoAd
30
Configure maximum capacity as 15 kg/30lb.
LoAd
30
LoAd
60
Configure maximum capacity as 30 kg/60lb.
LoAd
50
LoAd
100
Configure maximum capacity as 50 kg/100lb.