750000-INS-LAB-RevA08
13
s
Caution: Do not use corrosive or powdered cleansers
to clean the pump.
s
Caution: Do not immerse or soak the pump.
s
Caution: Do not clean with phenolic products.
Mattress
: Wipe the mattress unit with a clean, damp cloth
and mild detergent. The mattress may also be cleaned us-
ing a 10% solution of sodium hypochlorite (bleach) diluted
in water. Air dry all parts thoroughly before use.
s
Caution: Do not clean with phenolic products.
s
Caution: After cleaning, air dry the mattress without
direct exposure to sunlight.
general maIntenanCe
1. Check main power cord and plug for abrasion or exces-
sive wear.
2. Check mattress cover for signs of wear or damage. En-
sure mattress cover and tubes are connected correctly.
3. Check airflow from the air hose connector. The airflow
should alternate between each connector every half-
cycle time if it’s in alternating mode.
4. Check the air hoses for any kink or break. For replace-
ment, please contact your GF equipment provider.
fuse replacement
tool needed: small screwdriver
1. If you suspect a blown fuse, discon-
nect the plug from the wall outlet
immediately.
2. Use a small screwdriver to remove
the cover of the fuse holder (fuse lo-
cations are shown in picture at right).
3. Insert a new fuse of the correct rating (T1A/250V, VDE
approved).
4. Replace the fuse holder cover.