JB02020-INS-LAB-RevA20 • Fingertip Oximeter User Manual
13
Correct Measurement Procedures
• Ensure finger is correctly positioned in sensor.
• Relax; do not move; do not move finger during measurement.
• Do not place wet finger in sensor.
• Do not place the device on a limb with with a blood pressure
cuff applied.
• Do not place the device on a limb receiving venous infusion.
• Nail polish or paint may affect the measuring accuracy, and
too long a fingernail on measurement finger may cause
failure of measurement or inaccurate result.
• Vigorous exercise and electrosurgical device interference
may affect the measuring accuracy.
• Existence of high intensive light sources, such as
fluorescence light, ruby lamp, infrared heater or strong
sunshine, etc. may cause inaccuracy of measurement result.
Place an opaque cover on the sensor or change location.
• If the first measurement displays irregular waveform, then
the measurement is probably inaccurate; wait or restart.
Содержание John Bunn JB02020
Страница 25: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 25 Table 1 continued...
Страница 26: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 26 Table 2 continued...
Страница 27: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 27 Table 2 continued...
Страница 28: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 28 Table 2 continued...
Страница 29: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 29 Table 3 continued...
Страница 30: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 30 Table 3 continued...
Страница 31: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 31 Table 3 continued...
Страница 32: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 32 Table 3 continued...
Страница 33: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 33 Table 3 continued...
Страница 34: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 34 Table 4 continued...
Страница 35: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 35 Table 4 continued...