background image

VIC429-INS-LAB-RevA21 13

3.4  TOP AND BACK ADJUSTMENT

WARNING: PERSONAL INJURY HAZARD:

 WARNING: When the VIC is not in use, do 

not leave it in a recline position.

 WARNING: At no time should the patient 

be permitted to enter or exit from the 

ends of the VIC when in partial or total 

recline position.

 WARNING: When transporting patient, the 

patient should be placed in a fully upright 

position with chair in low height.

 WARNING: The patient should never be 

transported when the VIC is in recline 

position.

 WARNING: The VIC has a warning label 

on the top of the base cover that states: 

When chair is in nontransport mode: 

patient must remain fully upright; patient 

must never be reclined; patient must 

never be left unattended.

3.4.1  Raising the Top

Press the pump pedal (pedal nearest the head 
end) to the floor (Fig. 3.4-1), then release the 
pressure on the pedal. Repeat this process 
until the top is raised to the desired height. 
Use smooth strokes while pumping pedal to 
ensure patient comfort.

3.4.2  Lowering the Top

Press down on the release pedal (pedal 
nearest the foot end) (Fig. 3.4-2) until the top 
is lowered to the desired height.

Figure 3.4-1

Figure 3.4-2

Содержание Hausted VIC429

Страница 1: ...1 VIC429 VIDEO IMAGING CHAIR OPERATING MANUAL READ THIS MANUAL BEFORE OPERATING YOUR VIC429 VIDEO IMAGING CHAIR SAVE THIS MANUAL FOR FUTURE USE THE MOST CURRENT VERSION OF THIS MANUAL CAN BE FOUND ONLINE AT WWW HAUSTED COM ...

Страница 2: ...ON INSTRUCTIONS 9 3 1 3 VIC429 FEATURES ILLUSTRATED 10 3 2 VIC429 SPECIFICATIONS 11 3 3 BRAKING AND STEERING OPERATION 12 3 3 1 APPLYING THE BRAKES 12 3 3 2 RELEASING THE BRAKES 12 3 3 3 APPLYING THE STEERING LOCK 12 3 3 4 RELEASING THE STEERING LOCK 12 3 4 TOP AND BACK ADJUSTMENT 13 3 4 1 RAISING THE TOP 13 3 4 2 LOWERING THE TOP 13 3 4 3 RECLINING THE CHAIR BACK 14 3 4 4 RETURNING THE CHAIR BACK...

Страница 3: ...MS 20 TRANSFER OF PATIENT FROM VIC TO A SECONDARY DEVICE 20 3 10 PERFORMING PATIENT TRANSFER 21 PATIENT TRANSFER 21 PATIENT TRANSFER FROM A SECONDARY DEVICE TO VIC 21 4 PREVENTIVE MAINTENANCE FOR THE USER 22 5 OPTIONAL ACCESSORIES 23 6 GF HEALTH PRODUCTS INC LIMITED WARRANTY FOR HAUSTED BRAND STRETCHERS AND CHAIRS 24 7 DISPOSAL AND KEY TO SYMBOLS 26 DISPOSAL 26 KEY TO SYMBOLS 26 8 INDEX 27 COPYING...

Страница 4: ...patient treatment transport or recovery Service Information A thorough preventive maintenance program is essential to safe and proper unit operation This manual contains maintenance schedules and procedures which should be followed for satisfactory equipment performance We encourage you to contact GF Health Products Inc with maintenance concerns Advisory A listing of the safety precautions to be o...

Страница 5: ...d throughout this manual and their significance DANGER Indicates a potential hazard situation or unsafe practice that if not avoided will result in death or serious personal injury WARNING Indicates a potential hazard situation or unsafe practice that if not avoided could result in death or serious personal injury CAUTION Indicates a potential hazard situation or unsafe practice that if not avoide...

Страница 6: ...ient is seated in the chair ensure the side rails are up and the patient is secured with patient safety straps WARNING Patient entry egress and transfer from the chair should always be from the center side rail location with the side rail in the down position and the brakes locked WARNING At no time should the patient be permitted to enter or exit from the ends of the chair when in partial or tota...

Страница 7: ...Storage and Transport Temperature 10 C to 50 C Relative Humidity 20 to 90 30 C Atmospheric Pressure 700 to 1060 hPa UNPACKING INSTRUCTIONS IMPORTANT Follow each step in the order shown in these instructions Your Hausted equipment has been carefully packed at our manufacturing plant to ensure safe shipment to your medical facility There are several procedures you must follow to put your new equipme...

Страница 8: ...sed during transport easy to remove for radiological procedures Easy to operate rotating seat section Dual side hydraulic foot pedal controls height adjustment Four wheel brake and steer system Independent leg section Fold down height adjustable footrest Accessory wells two on each side Built In chart holder Removable push handles Four hook and loop safety straps for patient head chest waist and l...

Страница 9: ...and the patient is secured with patient safety straps G WARNING Patient entry egress and transfer from the chair should always be from the center side rail location with the side rail in the down position and the brakes locked H WARNING At no time should the patient be permitted to enter or exit from the ends of the VIC when in partial or total recline position I WARNING When transporting patient ...

Страница 10: ...RAP MOUNTING RODS See WARNING F REMOVABLE PUSHBAR See WARNING I GREEN PNEUMATIC ASSIST BACK RECLINE LEVER See WARNINGS B F H and I See Section 3 4 3 PATIENT ENTRY See WARNINGS G and H See Section 3 3 HINGING RAIL LATCH TOP ROTATION WARNING LABEL See WARNING K see Section 3 8 1 RED BRAKE LOCK PEDAL See WARNINGS G J See Section 3 3 HYDRAULIC HEIGHT ADJUSTMENT PEDAL See Section 3 4 INDEPENDENT LEG SE...

Страница 11: ...x 70 5 55 9 x 179 1 Unit Height 58 5 to 66 5 148 6 to 168 9 24 to 31 5 61 0 to 80 0 Back Section Width 14 35 6 14 35 6 Caster Diameter 5 12 7 5 12 7 Pad Multi Density Sectional Thickness 3 7 6 3 7 6 VIC429 Specification lb kg Product Weight 170 77 Shipping Weight 494 224 Maximum Weight Capacity Evenly Distributed 324 147 VIC429 Specification degree Trendelenburg Position 0 to 10 Seat Tilt 0 Back R...

Страница 12: ...ng 3 3 2 Releasing the Brakes To release the four wheel central braking system depress the green pedal on the patient s left at the VIC head end shown circled in red in Fig 3 3 2 until the pedal is in a horizontal position Fig 3 3 3 All four wheels will then rotate and swivel freely 3 3 3 Applying the Steering Lock From the VIC head end depress the green pedal downward into locked position Fig 3 3...

Страница 13: ...IC is in recline position WARNING The VIC has a warning label on the top of the base cover that states When chair is in nontransport mode patient must remain fully upright patient must never be reclined patient must never be left unattended 3 4 1 Raising the Top Press the pump pedal pedal nearest the head end to the floor Fig 3 4 1 then release the pressure on the pedal Repeat this process until t...

Страница 14: ...ht Position To return the back to upright position squeeze the recline lever and lift up on the back section Fig 3 4 4 3 4 5 Trendelenburg Position The VIC back has a Trendelenburg range of 0 horizontal position shown in Fig 3 4 5 to 10 Trendelenburg position To activate Trendelenburg position from any other position depress the green recline lever shown in Fig 3 4 3 Allow the back to lower into t...

Страница 15: ... while positioning the leg section Fig 3 5 2 Once desired position of the leg section is reached release the lever and the leg section will be locked in place Fig 3 5 3 When the leg section is independent it can be repositioned at any time by turning the lever clockwise Fig 3 5 3 s NOTICE Always support the leg section when turning the red lever 3 5 2 Returning to Chair Mode Hold and support the l...

Страница 16: ...sition With the pan in the retracted position pull out on the handles on either side of pan Fig 3 6 1 The footrest will drop into the lower position Fig 3 6 2 To place footrest in upper position Grasp the handles on both sides of the pan and tilt the pan up while lifting Fig 3 6 3 Once the pan is up all the way tilt the pan out until it locks into upper position Fig 3 6 4 To return the footrest to...

Страница 17: ...he rail Grasp the rail and lift it until the latching mechanism locks into place Fig 3 7 2 WARNING When patient is seated in the chair ensure the side rails are up and the patient is secured with restraint straps WARNING Rail release handle will not release unless force is applied laterally to outside top of rail WARNING The chair has a warning label located on both side rails indicating a pinch p...

Страница 18: ... 8 1 Mode Definitions Transport Mode The VIC is in transport mode when the green arrow by the seat lock knob is aligned with the green arrow on the base circled in red in Fig 3 8 1 and the seat is locked as shown in Figs 3 8 1 and 3 8 2 Nontransport Mode The VIC is in nontransport mode is when the seat has been locked 90 counterclockwise from transport mode as shown in Figs 3 8 3 and 3 8 4 WARNING...

Страница 19: ...e Lock the caster brakes depress the red pedal at the foot end of the unit on the patient s left circled in red in Fig 3 8 5 to approximately 45 Secure patient with safety straps Elevate chair to maximum height Fig 3 8 7 Pull out on the seat lock knob circled in red in Fig 3 8 6 to release lock Rotate chair counterclockwise until it locks into nontransport mode Fig 3 8 7 3 8 3 Transport Mode Later...

Страница 20: ...orted WARNING When chair is in nontransport mode patient must remain fully upright patient must never be reclined patient must never be left unattended Definition of Terms Used in the Following Instructions see Fig 3 10 1 below Transfer Side Terms Secondary Patient Device The transfer side will always be located on the attendant s right side when standing at the patient s foot end facing the patie...

Страница 21: ...g instructions on previous page Patient Transfer from a Secondary Device to VIC Follow the same procedure as in transferring from VIC to a secondary device on previous page except you must verify that the VIC is positioned so that the patient s abdomen is in line with the chair seat before attempting to transfer a patient to the VIC otherwise tipping may occur ...

Страница 22: ...t choice for vinyl Properly diluted quaternaries are also acceptable for vinyl Quaternary Isopropyl disinfectants ARE NOT recommended for vinyl Disinfectants for urethane products Standard Upholstery Quaternary disinfectants are recommended for urethane Quaternary Isopropyl disinfectants are recommended for urethane Phenolics SHOULD BE AVOIDED on urethane Disinfectants for all products All fabrics...

Страница 23: ...600 PEDIATRIC POSITIONER 36 X7 HSA063700 PEDIATRIC POSITIONER 48 X 10 HSA063800 PEDIATRIC POSITIONER 60 X 13 HSA0658 FOAM WEDGE PATIENT POSITIONING HSA071600 X WIDE SAFETY STRAPS WITH HOOK AND LOOP HSA071700 UNIVERSAL HEAD SAFETY STRAP HSA0721 PUSHBAR REMOVABLE STAINLESS STEEL COMES STANDARD HSA4002 VIC BACK EXTENSION ASSEMBLY VIC429 ONLY MFG AFTER 4 98 H000018 TELESCOPING I V ROD 27 TO 50 2 SECTI...

Страница 24: ...ment to the original customer unless there is an expiration date on the component in which case the warranty shall expire on the earlier of warranty period or the expiration date OBTAINING WARRANTY SERVICE Obtaining Warranty Service Inside the U S A GF Customer Service Representative must authorize warranty service Please contact the GF Customer Service department by calling 1 770 368 4700 or by e...

Страница 25: ...ITY AND THE LIABILITY OF GF S SUPPLIERS SHALL BE LIMITED TO THE FULLEST EXTENT PERMITTED BY LAW The warranties contained herein together with GF s Terms and Conditions found on GF s website and which may be updated from time to time contain all the representations and warranties with respect to the subject matter of this document and supersede all prior negotiations agreements and understandings w...

Страница 26: ...e and fiber content to aid sorting of plastic parts Product Metal Scrap Cable Scrap Electronic Scrap Plastic Recycling or Combustion VIC429 X X None X KEY TO SYMBOLS The following symbols are used on Hausted product labels Protective Earth Manufacturer Earth Ground Keep Dry General Warning Sign Fragile Handle with Care CE Mark Electrical and Electronic Equipment ETL Consult Instructions for Use EC...

Страница 27: ...Limited warranty 24 N Nontransport mode 18 Nontransport mode enter 19 NOTICE statement significance 5 O Operating instructions 8 P Patient transfer instructions 20 Patient transfer performing 21 Preventive maintenance for the user 22 R Rail lower 17 Rail raise 17 Rail repositioning 17 Rails hinging 17 S Safety statements significance of 5 Service information 4 Steering lock applying 12 Steering lo...

Страница 28: ...www grahamfield com 2021 GF Health Products Inc All Rights Reserved Hausted is a trademark of GF Health Products Inc One Graham Field Way Atlanta GA 30340 3140 GF Health Products Inc is an ISO 13485 2016 MDSAP Certified Company www hausted com www grahamfield com Manufactured by GF Health Products Inc One Graham Field Way Atlanta GA 30340 3140 ...