FR587WD-INS-LAB-RevB18 13

Adjusting Cable Tension 2

Adjusting Cable Tension 1

GAS SPRING TROUBLESHOOTING GUIDE

Info: gas springs have been pre-adjusted at factory

The gas spring on your recliner provides counterbalancing forces, and is self-locking for 
infinite, precise positioning. Unlocking the gas spring is controlled by actuating the lever 
located on the padded push handle on the back of the recliner. When the lever is actuated 
it pulls on the cable which in turn pulls on a hinge, depressing a button at the end of the 
gas spring rod. If the gas spring does not unlock when the lever is actuated or if it remains 
permanently unlocked when the lever is actuated, the cable must be adjusted.

Adjusting Cable Tension

1.  Remove the one screw and cable clip which secure the 

cable to the upholstered back. Remove the gas spring 
from the recliner by disassembling the two ring cotter 
pins and the two clevis pins that secure the gas spring to 
the back / frame.

2.  Lower the back until it rests on the stop bars on the 

recliner frame. Loosen the lock nut on the gas spring.

3.  Hold the hinge assembly while turning the gas spring 

until the cable just becomes taut. Back off the gas spring 
an additional half turn.

Info: If the gas spring was not unlocking when the 
lever was actuated, the gas spring must be turned 
clockwise. If the gas spring was not locking when the 
lever was released, the gas spring must be turned 
counter-clockwise.

4.  Retighten the lock nut.
5.  Reassemble the gas spring to the back of the recliner by 

replacing the clevis pins, bridge pins, and re-attach cable 
to back of recliner with the screw and cable clip.

6.  Actuate the recliner in accordance with the operating 

instructions to verify correct adjustment.

Содержание Lumex FR587WD Series

Страница 1: ...e with Drop Arm with Heat and Massage Assembly and Operating Instructions IMPORTANT READ THIS MANUAL BEFORE OPERATING YOUR LUMEX FR587WD FR587WDH DELUXE EXTRA WIDE CLINICAL CARE RECLINER SAVE THIS MAN...

Страница 2: ...RECLINE POSITION 10 TRENDELENBURG POSITION ALSO KNOWN AS SHOCK POSITION 11 DROP ARM OPERATION 12 LOWER DROP ARM 12 RAISE DROP ARM 12 GAS SPRING TROUBLESHOOTING GUIDE 13 ADJUSTING CABLE TENSION 13 HEAT...

Страница 3: ...nts used throughout this manual and their significance WARNING Indicates a potential hazard situation or unsafe practice that if not avoided could result in death or serious personal injury CAUTION In...

Страница 4: ...ous condition which may result in serious injury and will void the warranty WARNING This product can expose you to chemicals including Di 2 ethylhexyl phthalate DEHP which is known to the State of Cal...

Страница 5: ...lever total locking caster total of 3 on recliner black locking tab rear directional swivel locking caster right rear leg only green locking tab Series FR587WD Deluxe Clinical Care Recliner Extra Wide...

Страница 6: ...ck from the sides and place it horizontally with the upright back tubes resting on the recliner frame 6 Align the holes on the upright back tubes of the upholstered back with the holes on the seat fra...

Страница 7: ...frame crossbrace While holding the back in position with one hand secure both gas spring hinges to the U brackets by placing two 5 16 diameter clevis pins through the U brackets and hinges with the o...

Страница 8: ...te the total lock simply press down on the BLACK actuator tab until it locks To release the total lock press on the raised portion of the actuator tab until the lever is back in the normal operating p...

Страница 9: ...ngaged Lower Table Lift the side table up slightly until tabs disengage slots and pull table away from armrest Lower side table to self storing position SIDE PANEL OPERATION For easy cleaning and acce...

Страница 10: ...n either side of the recliner grasp the armrest with one hand and the push handle with the other hand Push back on push handle to recline To return grasp push handle and pull forward to raise the recl...

Страница 11: ...d be sitting in the recliner Trendelenburg Position Also Known as Shock Position Info This position can only be achieved by a caregiver Standing at the back of the recliner grasp the push handle with...

Страница 12: ...ipment are locked are at the same height in order to reduce the amount of lifting required to transfer the patient and are as close to one another as possible Failure to heed this warning could result...

Страница 13: ...which secure the cable to the upholstered back Remove the gas spring from the recliner by disassembling the two ring cotter pins and the two clevis pins that secure the gas spring to the back frame 2...

Страница 14: ...ubstitute parts Use only Lumex replacement parts The use of non Lumex replacement parts could cause personal injury property damage and void the warranty WARNING GF Health Products Inc assumes no resp...

Страница 15: ...nts as shown at right WARNING DO NOT place your hands in the path of the moving recliner mechanism Moving parts could cause personal injury 4 Use a 1 4 hex wrench to remove the two hex head screws tha...

Страница 16: ...the junction box cover replace the two hex head screws that secure the junction box cover see picture of junction box at top of previous page Use a 1 4 hex wrench to secure the screws 15 Gather loose...

Страница 17: ...g times a HEAT FUNCTION After one hour of continuous use at the same level b MASSAGE FUNCTION After fifteen minutes of continuous use at the same level c To restart the Heat and Massage Unit follow th...

Страница 18: ...he unit through all functions b If the switch seat symbol is rotated or there is evidence of damage to the box and cable notify Graham Field so that a switch box and cable can be sent for replacement...

Страница 19: ...ventilated area and keep it away from fire flame and recliner occupant Waxing and refinishing improves wearability Use any quality paste wax A diluted solution of 10 or less of bleach or Clorox may be...

Страница 20: ...registered disinfectants if available and use them in accordance with the manufacturer s instructions Do not use high level disinfectants liquid chemical sterilants for disinfection of either noncrit...

Страница 21: ...tive uses of mops cloths and solutions Prepare cleaning solutions daily or as needed and replace with fresh solution frequently according to facility policies and procedures Change the mop head at the...

Страница 22: ...3140 EC Authorized Representative CEpartner4U BV ESDOORNLAAN 13 3951 DB MAARN The Netherlands 31 0 6 516 536 26 www cepartner4u com TECHNICAL SPECIFICATIONS Overall Height 50 Overall Width 33 Overall...

Страница 23: ...ng but not limited to casters filters fuses gaskets lubricants and charts 4 Accessories or parts not provided by GF 5 Charges by anyone for adjustments repairs replacement parts installation or other...

Страница 24: ...rmation can be found on our website www grahamfield com 2012 GF Health Products Inc All Rights Reserved Graham Field and Lumex are trademarks of GF Health Products Inc GF Health Products Inc is an ISO...

Отзывы: