Pressing the S4 key will cycle between: Normal -> Swivel modes 
 
In the Swivel mode, the display will be rotated by 180°. 

 
 

Calibrate the reading 
 

You do not need to do the calibration if you just want to qualitatively detect radiation. 
However, if you have standard radiation sources with known µSv/h value, you may 
calibrate the instrument to get accurate readings. 
The factory default calibration value is for general in-door and out-door 
measurements. 
 
Calibration 

   

 

 
Select Calibrate form the Main Menu. There are three points provided for calibration. 
 

 

 
 
Each calibration point needs two values to be entered, the CPM value and the µSv/h value.  
The user can enter up to three calibration points from large to high count rates. 
 
Procedure: 

1.  Place the testing radiation source close to unit Geiger Tube side.  In Text or Graphic 

mode, wait for at least 3 minutes.  

2.  Note the CPM value. 
3.  Select Calibrate menu item on the device, then input both the CPM value you noted 

and the known µSv/h value of the calibration source. 

4.  Wait for the popup window to close to save the data for this calibration point. 
5.  Continue with step 1 with other calibrated radiation sources. 

 

 

 

 

 

Navigation keys: 

 

S1:  Back/Exit 
S2:  Increase the value by  10  CPM  or 0.1uSv/h respectively 
S3:  Decrease the value by  1  CPM  or 0.01uSv/h respectively 
S4:  Toggle input data type between CPM  and uSv/h 
 
 
 
 

Содержание GMC-300E Plus

Страница 1: ...GQ GMC 300E Plus Geiger Counter User Guide GQ Electronics LLC Revision 2 12 Oct 2017...

Страница 2: ...are 3 15 Re 1 01 Feb 2014 GQ Electronics LLC Revised for firmware 3 18 Re 2 00 Jan 2015 GQ Electronics LLC Revised for firmware 4 00 Re 2 10 Sep 2016 GQ Electronics LLC Content Revised Re 2 11 Sep 201...

Страница 3: ...ata Saving Setting 14 Add Note or Add Location for data saving 15 Note Location Input 15 History data searching 15 History data menu 15 Erase Saved History Data 16 Swivel Display Setting 16 Calibrate...

Страница 4: ...4 Tube recovery time 23 Extend battery operating time 23 Third party software developers 24...

Страница 5: ...r Counter software and the GQ Geiger Counter Data Viewer software The GQ GMC 300E Plus Soft Geiger Counter software s appearance is an exact copy of the GMC 300E Plus hardware human interface You can...

Страница 6: ...300E Plus main unit 2 USB cable 3 Quick start guide 4 Download link for SoftGeigerCounter software and USB driver for Windows 32 and 64 bits see Quick start guide For USB driver installation on Windo...

Страница 7: ...30 second if no key is being pressed during this period 4 Now the unit is ready to use You should see the background CPM rate in absence of a radioactive source Note There is a transparent protective...

Страница 8: ...e manger you should see which COM port is assigned to the GMC 300E Plus See the example below Windows assigned COM3 to the GMC 300E PLUS unit in this example For the Data Viewer or the data logger sof...

Страница 9: ...m upwards 4 In the menu mode while a popup message box is opened the S2 key changes the value by cycling through the predefined values S3 key 1 In text mode pressing the S3 key will change the display...

Страница 10: ...pup Window has timed out after 3 seconds if no key has been pressed Graphic User Interface GUI Graphic Icons Graphic ZOOM factors 0 1 0 2 0 3 1 0 2 0 3 0 4 0 5 0 AUTO Battery Status Data saving enable...

Страница 11: ...11 Average CPM of a specific time frame Maximum CPM since power on...

Страница 12: ...d time Current radiation level reading in CPM Current radiation level in Sv h or mR h format Total count and elapsed time since power on Press S2 for toggling displayed info Maximum CPM since power on...

Страница 13: ...ction S3 Sv h mR h CPM selection S4 Menu NOTE Large Font Mode is available in firmware Ver 2 20 or later only Menu Display Main Menu Navigation keys S1 Back S2 Up S3 Down S4 Select Enter User Option N...

Страница 14: ...Back Exit S2 Increase value by 1 S3 Decrease value by 1 S4 Select setting mode between Year Month Date Hour Minute and Second Each key press will change the mode to be set NOTE Setting of Date and Tim...

Страница 15: ...into the history data to identify the measurement or location This note location information will be saved together with the radiation data Note Location Input The cursor indicates the position of cur...

Страница 16: ...veraging mode Averaging time period is one minute Navigation keys for both data search modes S1 Back Exit S2 Select searching mode Previous Day Previous Hour Previous Minute Previous Note Location S3...

Страница 17: ...eds two values to be entered the CPM value and the Sv h value The user can enter up to three calibration points from large to high count rates Procedure 1 Place the testing radiation source close to u...

Страница 18: ...item and change the popup Windows message to Yes Wait for popup message window s timeout to confirm the reset request Any other key press will cancel the reset request It may take up to 30 seconds to...

Страница 19: ...be turned off if it has been set to ON The LCD back light will be off during the power saving mode Pressing any key will active the LCD backlighting again LED Indicator GMC 300E PLUS has a LED indicat...

Страница 20: ...arted after a scheduled time period passed This is useful if a testing target is not ready immediately When it started without a schedule the Timed Count will be started immediately once the timed cou...

Страница 21: ...to input data type between Day Hour Minute and Second Start a Timed Count If no Schedule set then the Timed Count starts immediately after the duration set If the Schedule is se then the Timed Count w...

Страница 22: ...22 About This menu item is used to gather information about the instruments model number firmware revision and serial number Model information Firmware version Unit serial number...

Страница 23: ...ort The USB port is the standard mini USB port It is used for data communication external power and the battery charger connection Data collection time The GMC 300E PLUS radiation data is collected co...

Страница 24: ...heir own software based on the published GQ RFC1201 protocol We encourage you to share your generic software with other users Please contact support gqelectronicsllc com if you have any questions You...

Отзывы: