Gossen Foto- und Lichtmesstechnik GmbH
27
8 Setting and measuring correction values
and extension factors
8.1 Setting correction values
- By simultaneously pressing both menu buttons
and
you reach the function
- correction values.
- The last valid correction value appears in the
display.
- The desired correction value can be entered or
altered using the setting wheel.
The extension factor is shown in the right digital display,
and the correction value is indicated in stops. Input in
1/10 EV (small digits) in the range of ±9.9 exposure value
stop increments.
For correction values which extend the exposure a
”
-
"
appears in front of the number.
Example:
-3.1 stops correspond to the extension factor 8.6.
For corrections which shorten the exposure, only the
left display appears as EV difference in stop
increments. By pressing one of the menu buttons
( or ) the correction value is stored in the memory of
the STARLITE 2.
The symbol
appears in the
display.
The correction value is automatically taken into account
in all measuring functions (except for the photometry).