GMC-I Messtechnik GmbH

3

Standard Equipment

1 test adapter (article number Z745V)
1 measurement cable red, 1 measurement cable black, 
2 measurement cables blue, 
2 alligator clips blue for contacting the welding electrodes, 
1 set of operating instructions German/English

Applications

In conjuction with a multimeter or a SECUTEST... test instrument, 
the 

SECULOAD

 test adapter is designed for testing welding equip-

ment (subsequenly referred to as device under test or DUT) in 
accordance with standard 

EN 60974-4:2007

. According to this 

standard, the peak values of the off-load voltage may not exceed 
the limit values for any possible setting.

Attention!

 The peak value of the open-circuit voltage is determined 

in the 

SECULOAD

 by means of a peak value rectifier with very fast 

diodes. As a result, the actual peak value of the open-circuit volt-
age is also issued for pulsed voltage sources with clock rates in 
the range of several 10 kHz, based upon the filter stipulated in the 
standard. For some arc welding equipment, the issued voltage 
value may exceed the admissible limit value due to the fact that 
such high-frequency clock pulses are not fully taken into account 
by the manufacturers of the welding equipment.

Special case: Welding instruments with an open-circuit voltage of less 
than 30 V *

In this case, the 

SECULOAD

 does not provides for any load control. 

Measurements are performed with a load resistance of 5 k

Ω

. For 

such low voltages, load control can be neglected, so that the 
measuring result is shown on the display even without pressing 
the 

START 

key.

* For the purpose of verification, perform a direct voltage measurement of 

your welding unit by using a multimeter or a SECUTEST... test instru-
ment (without having the test adapter connected in this case).

Special case: Welding instruments which deactivate open-circuit voltage

In this case, a certain key at the welding instrument must be 
pressed consistently in order to prevent deactivation. Please refer 
to the operating instructions of the welding instrument or contact 
the manufacturer for details on such cases.

Safety Instructions

This instrument fulfills the requirements of the applicable Euro-
pean and national EC guidelines. We confirm this with the CE 
marking. The relevant declaration of conformity can be obtained 
from GMC-I Messtechnik GmbH. The test adapter is manufac-
tured and tested in accordance with safety regulations 
IEC 61010-1/EN 61010-1/VDE 0411-1.
When used for its intended purpose, the safety of the operator, as 
well as that of the instrument, is assured.

Please make sure that the output of the DUT is de-energized for connect-
ing it to or disconnecting it from the test adapter.
Before using the test adapter, the off-load voltage of the welding equip-
ment must be checked. Input voltages at the test adapter of more than 
130 V are not permissible and result in the destruction of the test adapter. 
In this case, the welding equipment must be repaired prior to testing it 
with the test adapter.

In order to protect the internal resistor against overloading 
between the red and black socket connectors of the test adapter, 
these two 

socket connectors may not be short-circuited

. Wait until the 

measured voltage is no longer applied and/or the respective 
capacitors are discharged (this is performed automatically during 
operation with a SECUTEST... test instrument).
If the test adapter should nevertheless be overloaded once, volt-
age measurements are no longer possible. You are kindly 
requested to send the test adapter to GMC-I Service GmbH for 
repair in this case.

The test adapter may not be used

for testing plasma jet welding equipment

in the event of input voltages of more than 130 V

in the event of any visible external damage

if it no longer functions properly

Meaning of Symbols on the Instrument

Warning concerning a source of danger
(attention: observe documentation!)

CAT II

Measurement category II device

Continuous, doubled or reinforced insulation

Indicates EC conformity

This device may not be disposed of with the trash. Fur-
ther information regarding the WEEE mark can be 
accessed on the Internet at www.gossen-
metrawatt.com by entering the search term ’WEEE’

Carefully read the operating instructions before use, in particular the ope-
rating instructions included with the test instrument with which you in-
tend to use the adapter.

Test Sequences and Individual Measurements with the Test Adapter

The following test instruments include test sequences for welding 
instruments in accordance with EN 60974-4:

SECUTEST S2N+w

SECUTEST BASE(10)

SECUTEST PRO (or Feature I01)

SECUSTAR FM+

Alternatively, individual measurements, i.e. mere voltage tests can 
be performed in combination with the test adapter:

SECUTEST S2N+(10)

SECUTEST SIII+

Multimeter 

with min/max value memory

Connection of SECUTEST S2N+w / SECUTEST S2N+(10) 

Connect the welding current outputs with the input sockets of 
the test adapter (blue sockets).

Connect the test instrument (via test sockets 2 and 3) with the 
output sockets (+ and –) of the test adapter.

Connection of SECUTEST S2N+w / SECUTEST S2N+(10) 

GB

!

3

2

SECU

LOAD

blue

red

– 

black

Welding Equipment

Measurement of US
(Peak Value)

3

2

SECU

LOAD

blue

red

– 

black

Welding Equipment

Measurement of UR
(RMS Value)

Содержание SECULOAD Z745V

Страница 1: ...mbH angefordert werden Der Prüfadapter ist entsprechend den Sicherheitsbestimmungen IEC 61010 1 EN 61010 1 VDE 0411 1 gebaut und geprüft Bei bestimmungsgemäßer Verwendung ist die Sicherheit von Anwender und Gerät gewährleistet Versichern Sie sich dass der Ausgang des Prüflings zum Anschließen an den Prüfadapter oder zum Trennen von diesem jeweils spannungsfrei geschaltet ist Vor dem Einsatz des Pr...

Страница 2: ...tel Einzelmessungen unter Wechsel Gleichspannungsmessung beschrieben SECUSTAR FM Hier besteht keine Anschlussmöglichkeit für den Prüfadapter Die Spannungsmessung erfolgt mithilfe eines Multimeters Akti vieren Sie beim Einsatz eines Multimeters dessen Min Max Wertspeicherung SECUTEST PRO bzw Merkmal I01 Der Anschluss erfolgt hier wie bei Anschluss eines Multimeters über die Eingangsbuchsen COM und ...

Страница 3: ... as that of the instrument is assured Please make sure that the output of the DUT is de energized for connect ing it to or disconnecting it from the test adapter Before using the test adapter the off load voltage of the welding equip ment must be checked Input voltages at the test adapter of more than 130 V are not permissible and result in the destruction of the test adapter In this case the weld...

Страница 4: ...med in switch position Menu under parameter UAC DC as described in chapter Individual Measurements sub section Alternating Direct Voltage of the operating instructions for the test instrument SECUSTAR FM There is no connection provided for the test adapter Voltage is measured with a multimeter When using a multime ter activate its min max value memory SECUTEST PRO or Feature I01 Connection is esta...

Отзывы: