GMC-I Messtechnik GmbH
77
Meaning of Symbols in the User Interface – Test Sequence
Sym-
bol
Softkey Variants, Test Sequence
Test for Protection Class I Devices
Exposed, conductive parts are connected to the protec-
tive conductor so that they’re not charged with voltage if
the basic insulation should fail.
Test for Protection Class II Devices
These devices are equipped with double insulation or rein-
forced insulation.
Test for Protection Class III Devices
These devices are supplied with safety extra-low voltage
(SELV). Beyond this, no voltages are generated which
exceed SELV.
Type B applied parts (body) (see also section 14.7.2)
Type BF applied parts (body float) (see also section
14.7.2)
Type CF applied parts (cardiac float) (see also section
14.7.2)
“Applied part type” key: A group of applied parts can be
assigned to a type (B, BF or CF).
Prerequisite: The classification parameter and then the
applied parts menu have been accessed, and a group has
been created via the “Expand Group” key.
“Expand Group” key: Each time the key is activated, a fur-
ther applied part socket is assigned to the group. Prereq-
uisite: The classification parameter and then the applied
parts menu have been accessed.
“Reduce Group” key: Each time the key is activated, an
applied part socket is removed from the group. Prerequi-
site: The classification parameter and then the applied
parts menu have been accessed.
“Next Group” key: Each time the key is activated, the input
frame jumps to a position which has not yet been
assigned or to an already assigned position, depending
on whether you want to create a new group or edit an
existing group.
Prerequisites: The classification parameter and then the
applied parts menu have been accessed. At least one
group has already been created.
Configure sequence parameters (see page 83)
Set classification parameters
Assess visual inspection or function test with
OK
or
not OK
(toggle key)
Enter a comment, e.g. for the visual inspection or function
test
Continue test, next test step in the test sequence
Stop
continuous measurement
, next test in test sequence
Accept changed parameter,
return to memory view
Stop test sequence
– Repeat inspection (if it has been failed).
– Repeat test step
– Skip inspection test step
– Skip individual tests within the test sequence
This option can be enabled for the user in SETUP under
“Auto Measurements”.
Start evaluation – record measured value. Each time this
softkey is pressed, an additional measured value is saved
and the number is increased by one.
Start evaluation sequence during a
continuous measurement
.
The number blinks.
Record measured value during the evaluation sequence of a
continuous measurement
.
Repeat measured value recording
Delete measured value
Display measured values
Display details from the results list
Hide details from the results list
The ID number to which the measurement(s) will be stored
can be entered here.
Valid measured values have been obtained for a test
sequence. This measurement can be saved.
Save measurement data as (with display of directory path
/ ID or new entry of an ID other than the preselected one)
Transmit measurement data to a PC, e.g. in order to save
them to
IZYTRONIQ
report generating software (push-print
function) – refer to
IZYTRONIQ
online help for a description
Read-out of a complete test report at the end of a test
sequence
Read-out of a summarized test report at the end of a test
sequence
Read-out of all failed test steps instead of a test report at
the end of the test sequence
Sym-
bol
Softkey Variants, Test Sequence