GMC-I Messtechnik GmbH
83
Classification Parameter – IEC 60974-4
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
Sequence Parameters
The default test sequences can be adapted to your applica-
tion or test standard via the sequence parameters. The entered
sequence parameter settings are only valid for the currently
selected switch position (A1 to A9) and are retained until they are
changed. Not all of the parameters are relevant, depending on the
selected DUT classification (protection class etc.).
Parameter
Setting Options / Meaning
1/2
Standard
VDE 0701-0702, see table below
VDE 0701-0702-EDV, see table above
IEC 60601, 2
nd
edition, see table above
EC 60601, 3
rd
edition, see table above
EC 60601, 3
rd
edition GPA, see table above
IEC 62353, see table above
VDE 0701-0702-VLTG, see table above
VDE 0701-0702-PRCD, see previous table
IEC 60974-4
Protection class
1
PCI, PCII or PCI+II
Connection type
1
Test socket
Permanent connection
Adapter: AT16/32-DI adap.
Adapter: AT3-Adapter
2/2
Measurement type (MT)
1
Active
Voltage, rating plate
Voltage from rating plate, U(R) RMS
(RMS limit value, variably adjustable)
or open-circuit voltage U0 DC (limit value = 113 V DC)
Auto-detection
Connection & PC & MT
Connection & PC
Connection & MT
Connection only
PC & measurement type
Protection class only (PC)
Measurement type only (MT)
Disabled: No auto-detection:
All classification parameters such as connection, protection
class and measurement type must be entered manually.
Sequence Parameters
Meaning
Visual inspection (1)
Visual inspection (standard):
on: activate
off: deactivate
Visual inspection 2
(IEC 60974-4)
Visual inspection, function test, welding units
on: activate
off: deactivate
Function test
Function test:
on: activate
off: deactivate
Protective conductor resistance test
RPE
Protective conductor resistance test:
on: activate
off: deactivate
RPE IP
Protective conductor resistance at test socket:
Select test current IP:
±
200 mA= / 200 mA~ /
Feature G01: 10 A~ / feature G02: 25 A~
RPE IP permanent
connection
Protective conductor resistance with permanent connec-
tion:
Select test current IP:
±
200 mA= / 200 mA~ /
Feature G01: 10 A~ / feature G02: 25 A~
RPE as
Protective conductor resistance test:
Execute as individual or multiple measurement.
Multiple measurement: Repeat testing of various conduc-
tive parts as often as desired, in the event that it’s not
clear as to whether or not all accessible, conductive parts
are connected to each other or to the protective conduc-
tor.
RPE measurement
duration
Protective conductor resistance test:
A measurement duration within a range of 0 to 60
seconds can be entered here.
Insulation resistance test
RINS PC I
Insulation resistance tests for PCI:
on: activate
off: deactivate
RINS PC II
Insulation resistance tests for PCII:
on: activate
off: deactivate
RINS PC I and II
(VDE 0701-0702)
(IEC 60974-4)
Insulation resistance tests for PCI and II:
on: activate
off: deactivate
RINS at AP
Insulation resistance tests at applied parts:
on: activate
off: deactivate
Measurement duration
RINS probe
Insulation resistance test via probe:
A measurement duration within a range of 0 to 60
seconds can be entered here. Default setting: 3 s
Measurement duration
RINS AP
Insulation resistance tests at applied parts:
A measurement duration within a range of 0 to 60
seconds can be entered here. Default setting: 3 s
RINS pri./sec.
(VDE 0701-0702)
(IEC 60974-4)
Insulation resistance test between the primary and
secondary sides of PCIII DUTs
on: activate
off: deactivate
RINS PC II as
(VDE 0701-0702)
(IEC 60974-4)
Insulation resistance test:
Execute as individual or multiple measurement.
Multiple measurement: Insulation resistance is measured
between short-circuited mains terminals (L-N) and acces-
sible, conductive parts which can be contacted with test
probe P1 and are not connected to the housing, repeat as
often as desired.
Measurement duration
RINS PC II
Insulation resistance test:
A measurement duration within a range of 0 to 60 sec-
onds can be entered here. Default setting: 3 s
RINS sec./PE
(VDE 0701-0702)
(IEC 60974-4)
Insulation resistance test between the secondary side and
PE of PCIII DUTs:
on: activate
off: deactivate