14 GMC-I
Messtechnik
GmbH
4.9
Data Interfaces
4.9.1 RS 232 Service Interface
Figure 11
Diagram of Sub-miniature Plug for RS 232 Interface
Type
Service
Interface
Fieldbus Interface
Feature
F2
F1
F3/F4/F7/F8
Interface
RS 232
Profibus DP
CAN /
CANOpen
RS 485
Maximum number of
devices
1
32
100
32
Range of addresses
–
0 ... 126
0 ... 127
0 ... 254
Transmission speed
9.6 or 19.2
kBaud
9.6 kBaud ...
12 MBaud
10 kBaud ...
1 MBaud
9.6 or 19.2
kBaud
Protocol per
EN 60870
DIN 19245
part 3
IEC 1131
CANOpen
Modbus /
EN 60870/
HB-Therm/
DIN 19244
Connection
9-pin sub-
miniature plug
9-pin sub-
miniature plug
4-pole screw terminal
Connector Pin Assignments for
Sub-miniature Plug Connector for
RS 232 Service Interface
Establish
connection to
Connector Pin
Assignments at the PC
COM1 or COM2
Pin Number Designation
Description
Pin Number
Designation
2
TxD
Data output
2
RxD
3
RxD
Data input
3
TxD
5
GND
Signal ground
5
GND
1, 4, 6 ... 9
Not assigned
Socket
housing
Shield, con-
nected to con-
troller housing
9 8 7 6
5 4 3 2 1
9-pin sub-miniature plug connector at the controller