Test for nominal voltage 500 V and for the
„
Ω
“ range:
í
Set the rotary switch (6) to position „-Il-“.
The digital display shows the battery voltage at a simulated
load for the 500 V nominal voltage on the LCD display (5).
At the same time, the analog indicator shows the nominal
range of use of the battery voltage. You can quickly judge
the charging state of the battery set by the location of the
pointer.
The battery test for the nominal voltage 500V and for the
„
Ω
“ range is recognized by two superimposed horizontal
lines preceding the digital display.
Test for nominal voltages 250 V and 1000 V:
í
Set the rotary switch (6) to position „-I|-“ as described
above.
í
For the test of the nominal voltage 250 V, briefly press
the „RANGE
↓
“ pushbutton (4), or briefly press the
„RANGE
↑
“ pushbutton (3) for the test for the nominal
voltage 1000„RANGE
↑
“V.
The battery voltage is displayed in digital and analog form,
as described above.
For recognition, a horizontal line precedes the digital display
for 250 V nominal voltage and three superimposed lines for
1000 V nominal voltage.
If the battery test reveals that the battery voltage is below
the lower limit, „U LO“ is shown on the LCD display (5).
The battery test is automatically stopped after about
10 seconds. „HOLD“ is shown on the LCD display (5) in
addition to the digital display and the analog indication.
This presentation is maintained until the tester automatically
switches back to „standby“ after approximately 3 minutes,
until you press the measurement pushbutton (8) or until you
set the rotary switch (6) to another position.
You start a new battery test by pressing the measurement
pushbutton (8).