Ascertaining Average Temperature – AVG Function
This measuring function makes it possible to acquire and store the
average temperature value after scanning at least eight measuring
points (SCAN function) during a single measuring sequence.
➭
Repeatedly press the
MODE
key until
AVG
appears at the display.
➭
Pull and hold the trigger until all measuring points have been scanned.
The average temperature value is frozen at the auxiliary display after
the trigger is released.
Ascertaining Temperature Difference –
T Function
This measuring function makes it possible to acquire and store the
largest measured temperature difference (difference between lowest
and highest measured temperature value) after scanning several
measuring points (SCAN function).
➭
Repeatedly press the
MODE
key until
T
appears at the display.
➭
Pull and hold the trigger until all measuring points have been
scanned.
The temperature difference is frozen at the auxiliary display after the
trigger is released.
4.1.3 Memory Functions
Saving Measurement Data – DATA Function (blinking)
Up to 10 measured values can be saved to 10 memory locations:
DATA1 ... 9, A, where A is used as a hexadecimal character for the
10
th
memory location.
The current measured value, the maximum value (MAX), the minimum
value (MIN), the average value (AVG), maximum temperature difference
T or temperature measured by means of a thermocouple (TC) can be
stored to any given memory location.
Proceed as follows in order to save values to memory:
➭
Repeatedly press the
MODE
key until
DATA
blinks at the display.
➭
Select
DATA1
by pressing the
or the
key.
➭
Repeatedly press the
MODE
key until the desired measuring
function appears at the display.
➭
Start measurement by pulling the trigger.
➭
Save the measured value to memory during measurement by
pressing the
M
key. A brief acoustic signal acknowledges that the
value has been saved. The memory location number is increased
by one.