Gossen MetraWatt METRALINE RLO-CHECK Скачать руководство пользователя страница 4

 

RECALIBRATION TO 2.00 

Ω 

 



(The METRALINE RLO-CHECK is switched on.) Undo all 

connections (terminals and role-up cable).

 

 



Check to see if the battery icon indicates full charging (if not, 
replace the batteries with new ones, see section J). 



Locate the 0 

Ω

 short-circuit jumper 

(

M1).

 

 



Briefly press the « 

ϕ 

» key 

twice



Connect the 0 

Ω

 shunt 

(

M1) 

to the two terminals. 



Wait for approximately 3 seconds. 



Blue background, 0.00, two short acoustic signals 

followed by a continuous acoustic signal 

indicate that balancing has been 

successfully completed (the displayed 

value may fluctuate due to measuring 
tolerance, for example between 0.00 and 

0.07). 



Remove the 0 

Ω

 short-circuit jumper 

(

M1). 



Press and hold the « 

ϕ 

» key for at least 2 seconds in order to 

switch the device off. 



Locate the 2 

Ω

 calibrating shunt (M2) and a 4 mm flathead 

screwdriver. 



Loosen and remove the screw (J1), and remove the battery 
compartment lid (J2).

 

 



Locate a precision screwdriver with a maximum diameter of 

2 mm.

 

 



Press and hold the « 

ϕ 

» key until you hear a double acoustic 

signal in order to switch the device back on again.

 

 



Connect the 2 

Ω

 calibration shunt (M2) to the two terminals. 



Using a precision screwdriver with a max. diameter of 2 mm, 

press the key inside the housing to recalibrate (opening M3).

 

 



Blue background, 1.00, and the CAL icon

 indicate that 

recalibration to 2.00 

Ω 

has been successful.

 

 



Press and hold the « 

ϕ 

» key for at least 2 seconds in order to 

switch the device off. 



Remove the 2 

Ω

 calibration jumper 

(

M2). 



Replace the battery compartment lid (J2) and retighten the 
screw (J1). 



Press and hold the « 

ϕ 

» key until you hear a double acoustic 

signal in order to switch the device back on again. 



Balance resistance to your accessories in accordance with 

section E or G. 

 
The METRALINE RLO-CHECK is ready for measurement. 

Recalibration to 2 

Ω 

remains effective until the next 

recalibration, even if the battery is replaced and/or the 
METRALINE RLO-CHECK is turned off. 

 

 

Red background, E02 and CAL icon:

 

 

Recalibration was not successful

 

(

e.g. 

because the wrong calibration shunt was 

used etc.).  

Switch the

 

METRALINE RLO-CHECK off. 

 

Ω

 

 

Ω

 

0.0

 

E0

CAL 

M1 

 

 

CAL 

 

CAL 

CAL 

 

2.0

CAL 

Ω

 

M2 

 

 

J1 

J2 

M3 

 

Subject to change without notice • PDF version available on the Internet

 

 

 
 
 
 

 

 

Phone  +49 911 8602-111

 

GMC-I Messtechnik GmbH  

 

Fax 

+49 911 8602-777 

Südwestpark 

15 

   E-Mail 

[email protected] 

90449 Nürnberg 

Germany  

www.gossenmetrawatt.com 

Repair and Replacement Parts Service

 

Calibration Center and Rental Instrument Service

 

 

If required please contact:

 

 

 
GMC-I Service GmbH 
Service Center 
Beuthener Str. 41 

90471 Nürnberg 

 Germany 

Phone:  +49 911 817718-0 

Fax: 

+49 911 817718-253 

E-Mail  

[email protected]

 

www.gmci-service.com

 

 

This address is only valid in Germany.

 

Please contact our representatives or subsidiaries  

for service in other countries 

 

 

Product Support 

If required please contact: 
GMC-I Messtechnik GmbH 
Product Support Hotline 

Phone   +49 911 8602-0 
Fax:  

+49 911 8602-709 

E-Mail  

[email protected]

 

Содержание METRALINE RLO-CHECK

Страница 1: ...ed by a continuous acoustic signal indicate that balancing has been successfully completed the displayed value may fluctuate due to measuring tolerance for example between 0 00 and 0 07 The METRALINE RLO CHECK is ready for measurement The results of this balancing procedure are retained until balancing is conducted again even if the battery is replaced or the METRALINE RLO CHECK is turned off Red ...

Страница 2: ...s Approx 5 hours and 50 minutes for continuous measurement of volume resistance of an earthing conductor of less than 1 Ω with activated acoustic signal and new batteries The instrument is switched off automatically after five minutes of nonuse Conformity with standards EN IEC61010 1 2010 EN IEC61557 4 2007 Conformity with European directives 2011 65 EC RoHS directive 2006 95 EC low voltage direct...

Страница 3: ...r see I 0 5 A fuse protected Warning regarding interference voltage If interference voltage is applied to the protective conductors resistance measurement is disabled An intermittent acoustic signal is generated the color of the display is changed to red and the interference voltage value is displayed If a minus sign appears to the left of this value the interference voltage is direct voltage and ...

Страница 4: ...for at least 2 seconds in order to switch the device off Remove the 2 Ω calibration jumper M2 Replace the battery compartment lid J2 and retighten the screw J1 Press and hold the ϕ key until you hear a double acoustic signal in order to switch the device back on again Balance resistance to your accessories in accordance with section E or G The METRALINE RLO CHECK is ready for measurement Recalibra...

Отзывы: