7.7.2 Alternating Current and Frequency Measurement, Direct Connec
-
tion, A AC and Hz
➭
First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
➭
In accordance with the current or frequency to be measured,
turn the rotary switch to A~ or Hz.
➭
Select the desired measured quantity by briefly pressing the
FUNC | ENTER
multifunction key. Each time the key is pressed,
AC
TRMS
and Hz are alternately selected, and
switching is acknowledged with an acoustic signal.
➭
Safely connect the measuring instrument (without contact
resistance) in series to the power consumer as shown.
➭
Switch supply power to the measuring circuit back on (3).
➭
Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
➭
Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
➭
Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
COM
mA A
~
R
x
~
R
x
1
3
2
A~
Hz
A~
Hz
003
.
50
A
AC
TRMS
050
.
10
Hz
AC
8
.
8
.
8
.
8
.
8
!
I > 10 A
!
I > 16 A
Current
FUNC
ENTER
METRAHIT PM XTRA / PM TUNE
/
E
XTRA
METRAHIT EU PRO / E
SPECIAL
6 A / 10 A (16 A max. 30 s)
600
A / 6 mA
60 mA / 600 mA
6 A / 10 A (16 A max. 30 s)
Measuring Ranges:
METRAHIT PM TECH / E
TECH
60 mA / 600 mA
6 A / 10 A (16 A max. 30 s)
Current measurement
may only be performed
with installed batteries!