GMC-I Messtechnik GmbH
31
GB
Replacing the Batteries
Warning!
Use the included type-tested batteries only, as de-
scribed in the Characteristic Values.
➭
Set the instrument face down onto a flat working sur-
face, loosen the two screws at the back and lift off the
housing base, starting at the bottom. The housing top
and housing base are held together with the help of
snap hooks at the top front.
➭
Remove the batteries from the battery compartment.
➭
Insert four 1.5 V batteries into the battery compart-
ment, making sure that the plus and minus poles
match up with the provided polarity symbols.
➭
Important for reassembly: First set the housing base
onto the housing top and align accurately (see photo
below). Then press the two housing halves together,
first at the bottom front (a), and then at the top front (b).
➭
Secure the housing base with the two screws.
Warning!
The instrument may not be operated if the housing
base has not been properly installed and secured!
Disposing of Batteries
Dispose of old batteries properly, i.e. at a designated col-
lection point.
10.2 Fuses
Inspecting the Integrated Fuse F1
The fuse F1 is located at the measuring input. When the
fuse is defective, the measurements in the m
Ω
/
Ω
ranges
are faulty.
➭
Short circuit the „I+“ and „I–“ as well as the „U+“ and
„U–“ jacks.
Display < 0.4
Ω
: fuse OK.
Display „
0L
“ and „
openL
“:
fuse blown.
Attention!
!
If a fuse should blow, eliminate the cause of overload
before placing the instrument back into service!
(b)
(a)