8.4
MAVOPROBE LUX / UVA
– Illuminance, Class B / UV-A Irradiance
The high-precision MAVOPROBE LUX/UVA combination probe for non-destructive testing is classified
for class B illuminance in accordance with DIN 5032-7, DIN EN 13032-1 appendix B and
ISO/CIE 19476, and measurement of UV-A irradiance complies with requirements set forth in
DIN EN ISO 3059 and ASTM E2297 for measuring instruments for fluorescent penetrant and magnetic
particle testing.
Performance of the non-destructive materials testing system must be checked regularly to ensure
inspection quality and reliability. This inspection includes the intensity of the UV-A radiation as well as
illuminance. The viewing conditions for this test method are described in DIN EN ISO 3059, which
contains the minimum requirements for illuminance and UV-A irradiance and their measurement.
Calibration intervals according to the manufacturer’s specifications, but in any case at least once every
12 months, are also stipulated and must be verified with a calibration certificate.