background image

3-349-600-37  6/11.16 

 

 

 

 

FIG 1 

1. 

Jaws 

2. 

Conventional current direction 

3. 

Safety zone delimiter (Tactile Barrier): 

Do not reach beyond the safety collar!

 

4. 

Battery cover 

5. 

Battery cover screw 

6. 

Jaw trigger 

7. 

ON/OFF / Range switch 

8. 

Auto zero button 

9. 

LED 

 

 

4.  OPERATING INSTRUCTIONS 

4.1  Switch On

 

When  the  probe  is  switched  on,  the 

green  LED 

will 

illuminate.    The  LED  starts  flashing  when  the  battery 
voltage is too low for normal operation and warns the user 
that it requires changing.  This procedure is described in 
Section 5.

 

4.2  Auto Zero 

 

The output zero offset voltage of the probe may change 
due to thermal shifts and other environmental conditions.  
To null the output voltage depress the Auto Zero button.  
Ensure that the probe is away from the current carrying 
conductor whilst the probe is being nulled. 

4.3  Current Measurement

 

Switch on the probe and check that the LED is lit. 

Connect the output lead to an oscilloscope, multimeter or 
other measuring equipment. 

Zero the probe using the Auto Zero button. 

Clamp  the  jaws  of  the  probe  round  the  conductor 
ensuring a good contact between the closing faces of the 
jaws. 

Observe  and take measurements as  required.  Positive 
output  indicates  that  the  current  flow  is  in  the  direction 
shown by the arrow on the probe. 

4.4  Auto Power OFF 

In order to save battery life, the probe will automatically 
switch  itself  off  after  approximately  10  minutes.    To 
disable the Auto power off function, Switch Off the probe 
and Switch On whilst pressing the auto zero button. The 
red  LED  will  illuminate  and  the  probe  will  stay  On  until 
switched off again.  

4.5  Battery Replacement 

 

The  green  or  red  LED  will  flash  when  the  minimum 
operating voltage is approached.  Refer to Fig.1.  Use the 
following procedure.

 

Unclamp  the probe from the conductor, turn it off using 
the

 On - Off

 switch and disconnect the output leads, from 

external equipment. 

Loosen  the  captive  screw  which  secures  the  battery 
cover.  Lift the cover through 30° and pull it clear of the 
probe body as shown in Fig 1.  Replace the battery and 
re-fit the battery cover and fasten the screw. 

Fit only Type 9 V PP3 Alkaline (MN 1604).

5.  SAFETY STANDARDS

 

EN 61010-1 
EN 61010-2-032 
 

300V Cat 

III 

Pollution Degree 2 

EMC Standards

 

EN 61326-2-2 
ROHS and WEEE compliant 
This  product  is  designed  to  be  safe  under  the 
following conditions: 
 

-  indoor use 

 

-  altitude up to 2000m 

 

-  temperature 0°C to +50°C 

 

-  maximum relative humidity 80% for 

 

  temperatures up to 31°C decreasing linearly 

 

  to 40% relative humidity at 50°C.

 

Use of the probe on 

uninsulated conductors

 is 

limited  to  300V  AC

RMS 

or  d.c.  and  frequencies 

below 1kHz. 

Safety in its use is the responsibility of the operator 
who  must  be  a  suitably  qualified  or  authorised 
person.  Ensure that your fingers are behind the 

protective  barrier

  see  FIG  1  when  using  the 

probe.    Always  inspect  the  probe  for  damage 
before use. 
 
To avoid electric shock, keep the probe clean and 
free of surface contamination. 
 
Use Isopropyl alcohol to clean the probe. 
 

6.  WARRANTY 

Your AC/DC current probe is guaranteed for two 
years from the date of purchase against defective 
material or workmanship. If the unit

 

fails during 

the warranty period, we shall at our discretion, 
repair or replace it with a new or reconditioned 
unit provided we are satisfied that the failure is 
due to defective material or workmanship. To 
make a claim under warranty, the probe should 
be returned to us, postage prepaid, with a 
description of the defect. The use of a battery or 
external power supply, other than that specified 
invalidates this warranty.

 

Goods  alleged  by  the  buyer  to  be  defective 
shall  not  form  the  subject  of  any  claim  for 
injury,  loss,  damage,  or  any  expense 
howsoever incurred whether arising directly or 
indirectly from such alleged defects other than 
death  or  personal  injury  resulting  from  the 
seller‘s negligence. 

 
No  condition  is  made  or  to  be  implied  nor  is  any 
warranty given or to be implied as to the life or wear 
of goods supplied or that they will be suitable for any 
particular  purpose  or  for  use  under  specific 
conditions,  notwithstanding  that  such  purpose  or 
conditions may be made known to the seller. 
 

7   PRODUCT SUPPORT 

If required please contact: 

GMC-I Messtechnik GmbH 
 

Product Support Hotline 

Phone: 

+49 911 86 02-0 

Fax:  

+49 911 86 02-7 09 

Email: 

[email protected]

 

 

 

8   REPAIR AND REPLACEMENT PARTS 
SERVICE DKD CALIBRATION CENTRE 
AND RENTAL INSTRUMENT SERVICE 

If required please contact: 

GMC-I Service GmbH 
Service Center 
Thomas-Mann-Strasse 20 
90471 Nürnberg, Germany 
 
Phone: 

+49 911 817718-0 

Fax:  

+49 911 817718-253 

E-mail: 

[email protected]

 

 
This  address  is  only  valid  in  Germany.  Please 
contact  our  representatives  or  subsidiaries  for 
service in other countries. 

 
 
 
 

CP30_CP330_CP1100  English Rev 1.4 

 

 

1

2

5

4

7

9

8

6

4

3

SAFETY WARNING 

Before removing the battery cover, 
make sure that the probe is remote 

from any live electrical circuit.

 

Содержание CP 30

Страница 1: ...h die Stromsensoren zur exakten Strommessung bis 30 A CP 30 bis 450 A CP 330 bis 1400 A CP 1800 bzw bis 1800 A CP 1800 Folgende Messeigenschaften sind verfügbar AC und DC Strommessung Hohe Genauigkeit 1 mA 50 mA 100 mA Auflösung Autom Abschalten Diese Eigenschaften machen es zum leistungsstarken Instrument zur Messung an Umrichtern getakteten Schaltnetzteilen industriellen Steuerungen und anderen ...

Страница 2: ...tive Luftfeuchte von 80 bis 31 C und einer sich linear verringernden relativen Luftfeuchte von 40 bei 50 C Der Einsatz des Stromsensors an nicht isolierten Leitern begrenzt sich auf 300 V ACeff oder DC bei Frequenzen unter 1 kHz Die Sicherheit im Gebrauch unterliegt der Verantwortung des Anwenders der eine entsprechend qualifizierte oder berechtigte Person sein muss Stellen Sie sicher dass Ihre Fi...

Страница 3: ... and other applications requiring current measurement and or waveform analysis 3 SPECIFICATION 3 1 Electrical data All accuracies stated at 23 C 1 C CP 30 Z201B Measuring range 0 to 30A DC or ACpk Current Range 30A Resolution 1mA Overload capacity 60s 500A Basic accuracy DC 300Hz 1 of rdg 2mA Temperature Coefficient 0 01 of reading C Frequency range DC to 20kHz 1dB CP 330 Z202B Measuring range 0 t...

Страница 4: ...titude up to 2000m temperature 0 C to 50 C maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 40 relative humidity at 50 C Use of the probe on uninsulated conductors is limited to 300V ACRMS or d c and frequencies below 1kHz Safety in its use is the responsibility of the operator who must be a suitably qualified or authorised person Ensure that your fingers are behind ...

Страница 5: ...3 349 600 37 6 11 16 ...

Отзывы: