Display for parameter setting Trigger Source OFF
Rotating points
indicate that data is
being stored to
memory.
L .o...... .G .G.
0 ... 99 %, FULL
(memory occupancy level)
If the power supply is interrupted during a recording session, the A2000 supplements the outstanding
samples after restarting the instrument:
•
A value of „0“ is entered for all measured quantities, except for energies (last meter reading)
•
If a trigger source has been selected, the beginning of power supply interruption is always consid-
ered to be a trigger.
•
If the trigger source has been set to „OFF“, the beginning of power supply interruption is recorded
in the time stamp of the last trigger. (Time stamp of the first trigger = start of recording)
•
If power supply interruption takes longer than the remaining storage rate, the current window is
completed and a new untriggered window is produced if a trigger source has been selected.
•
In the case of trigger source „OFF“, cyclical memory mode and a power supply interruption
which takes longer than the storage rate, the complete memory will be overwritten.
!