INTRODUCTION
17
Connecting GSB-02 Units
Background
The following will describe how to connect
scanner boxes to a GPT-9000 series safety tester
with ground bond test support. Up to 4 scanner
boxes can be connected. In the examples below
only 3 scanner boxes are connected.
When wiring the GSB-02 scanner boxes to the
safety tester or to each other, only the GB
wiring leads (GHT-109) should be used.
Front Panel
1.
Connect the SOURCE H and SOURCE L
terminals on the safety tester to the same
terminals on the 1
st
scanner box.
2.
Connect the SENSE H and SENSE L terminals
on the safety tester to the same terminals on the
1
st
scanner box.