How to Handle Calibration Objects
Sensor Calibration
Page 28 (40)
at
os
_cs
_r
ev
01_en
_r
ev
-c
2012
-09
-03
10.
Sensor Calibration
Calibration is a measuring process during which the measuring sys-
tem with the help of calibration objects is adjusted such that the di-
mensional consistency of the measuring system is ensured. During
this process, the software determines geometrical parameters, for ex-
ample position and orientation of each camera, based on the recorded
camera images. In addition, the image characteristics of the camera
lenses and the camera chips are determined.
Based on these settings, the ATOS software calculates from the points
of the calibration object in the 2D camera image their 3D coordinates.
For the ATOS COMPACT SCAN SENSORS, three different calibration
objects are used (panel with magnetic support, panel in the case, and
cross).
Generally, each measuring volume (MV) has its own calibration object.
The table in section 4.1 informs you about which calibration object is
required for your measuring volume.
Crosses are used for system types for mobile operation, because the
panels are too large for measuring volumes
≥
MV 250.
Panels with a magnetic support are suitable for the smallest measur-
ing volumes, e.g. MV 170 for camera position
SO
.
Calibration panel for MV 125
Calibration panel in the case for MV 300 Calibration cross for MV1200
If for a measuring volume you may use both, panel calibration and
cross calibration objects, please use the more accurate panel cali-
bration.
Carry out the calibration only with the calibration object intended for
your measuring volume.
10.1
How to Handle Calibration Objects
Always handle the calibration objects with utmost care and prevent
them from getting dirty and being scratched. Make sure you do not
touch the surface of the calibration object if possible. After each use,
duly accommodate the calibration objects.
Clean the surface of the calibration object using the enclosed micro-
fiber cloth from the standard tool box. Wipe the surface gently only.
Never exert any pressure.
Microfiber cloth for cleaning the calibration panels