Sensor Setup
8.1
General
In order to adjust the sensor, the complete system including the
ATOS software needs to be installed and the sensor must be
equipped with a set of lenses matching the measuring volume and
the camera position. For achieving all camera positions, three differ-
ent camera supports can be used.
Camera position
500
Camera position
300
Camera position
SO
For each measuring volume (MV), the set of lenses are factory-
preadjusted.
If you have the respective set of lenses for each measuring volume,
setting up the sensor is easy as you do not need to carry out all the
steps referring to adjusting the focus and aperture!
You always need to perform a complete sensor setup if, for example,
the corresponding set of lenses has never been adjusted before or is
decalibrated.
8.2
Measuring Volume
Use the table in section 4.1 to determine a measuring volume.
You create larger measuring volumes with camera positions
500 and
300
and smaller measuring volumes with camera position
SO
(see ta-
ble 4.2).
Please check whether the cameras are in the correct position. How to
change the camera positions is described in section 0.
8.3
Changing the Camera Positions (Step by Step)
In the following, we describe how to change the camera distance from
300
to
SO
.
When you are changing the distance from
SO
to
300
or when you
are changing to camera position
500
, please proceed corresponding-
ly. All camera supports have identical connecting elements.
The length of the camera connection cables are matched to the
camera support when the system is delivered.
If you received the support for camera position
500
or
SO
later, the
connection cables might not be long enough. In that case, contact
the GOM support or your local distributor.