1.0  INTRODUCTION

The XRS3 is a small, lightweight X-ray generator that operates on its own removable battery pack. The XRS3 is a 
pulsed X-ray device that produces X-ray pulses of very short duration (25 nanoseconds).  It produces a relatively 
low dose rate comparable to a 0.25 ma constant potential machine. The energy produced by the XRS3 is up to 
270kVp, which makes it possible to radiograph up to one (1) inch (2.54 cm) of steel.

XRS3 standard accessories are two keys, two battery packs, and one battery charger. Remote cable and carrying 
case are also common accessories. 

2.0   WARNINGS

The XRS3 is a pulsed X-ray generator that emits hazardous ionizing radiation when pulsing.  The XRS3 should only 
be 

operated

 by 

authorized personnel

 who are properly trained to safely operate the generator.  The XRS3 must 

be 

registered

 with proper authorities prior to use and should 

not

 be used to intentionally expose humans.      

Develop and closely follow a safe operating system for using the XRS3.  The safe operating system must ensure that 
no one is exposed to radiation above the permissible limits which are 2 mR (0.02 mSv) per hour for a member of 
the public.  

The safe operating system must ensure the XRS3 is used within federal and state guidelines. 

All operators and users of the XRS3 X-ray machine must wear a personal radiation monitoring device, such as 
a  TLD  (thermoluminescent  dosimeter),   lm  badge,  and/or  a  pocket  dosimeter 

consistent with  the appropriate 

federal, territorial or provincial standards

 (note: an electronic dosimeter will not detect the XRS3 radiation pulses).

Due to the short pulse width of the XRS3, survey meters of the Geiger-Mueller and scintillator type do not accurately 
detect the radiation emitted from the x-ray source.  

Survey meters should be of  the ionization type and should be used in the integration mode

.  

Survey meters must not 

be used in the rate mode because the XRS3 does not produce constant radiation.  The XRS3 produces very high 
rates of radiation for very short periods of time resulting in either unrealistically high readings or no readings for 
a survey meter in rate mode. 

The XRS3 has no explosion proof rating and should not be used in an explosive atmosphere. The Spark Gap 
is vented to the air and could be a source of ignition. 

STORAGE & USE:  Store and use the XRS3 in upright position.  

Upside down storage or use may cause premature 

failure including no output dose.   

The XRS3 produces high levels of radiation and must be operated by quali  ed personnel who have read the

WARNINGS and OPERATING INSTRUCTIONS sections of the manual before operating the device.

DUTY CYCLE WARNING.

  

The XRS3 is a light duty machine that is not made to pulse continuously.  

The 

maximum duty cycle for the XRS3 is 200 pulses every four minutes (3000 pulses per hour). Exceeding the 
duty cycle will shorten the life of the tube and head. In temperatures above 90 °F (32.22 °C) or continual 
use, rest 30 seconds after every 50 pulses and 4 minutes after every 200 pulses.

Содержание XRS-3

Страница 1: ...OPERATOR S MANUAL MARCH 2018 Golden Engineering Portable X ray Technology...

Страница 2: ...ERATING PRECAUTIONS 8 EXCLUSION ZONE 8 PULSE SETTINGS 8 6 0 NAVIGATING THE MENU 9 7 0 MAINTENANCE 11 DOSE MEASUREMENT 11 TUBE REPLACEMENT 11 8 0 TROUBLE SHOOTING 11 9 0 INSTRUCTIONS FOR SERVICE 12 REM...

Страница 3: ...dge and or a pocket dosimeter consistent with the appropriate federal territorial or provincial standards note an electronic dosimeter will not detect the XRS3 radiation pulses Due to the short pulse...

Страница 4: ...mount The identi cation label located on the bottom of the generator lists the manufacturer s name and address model number serial number weight volt amp and production date BATTERY PACK The standard...

Страница 5: ...igit LCD Home displays number of pulses selected and delay time before the rst pulse Home will also indicate if multiple pulse trains have been entered Arrows allow operator to scroll through various...

Страница 6: ...ABLE CONNECTOR Lemo K series ve pin connector located on the back of the control module beneath the battery receives the remote cable or imaging system cable POSITIVE REMOTE CABLE CONNECTOR BATTERY TE...

Страница 7: ...ulsed 7 As the High Voltage Switch is closed a high voltage transient of approximately 270 000 volts and 25 nanoseconds in duration is applied across the x ray tube generating x rays The closing of th...

Страница 8: ...7 5 0 OPERATING INSTRUCTIONS The following are basic operation instructions to take an X ray image using the XRS3 Certain applications may require modi cations to these basic procedures...

Страница 9: ...feet 3m behind the X ray unit and clear all personnel at least 10 feet 3m behind the unit or at least 100 feet 30m from the front of the unit before pulsing The exclusion zone below should be a contro...

Страница 10: ...pulse setting LCD ERROR MESSAGE Low Battery Battery voltage is too low to pulse the XRS3 Charge the battery PWM not running Could occur at the start of a pulse train or during a pulse train Power uni...

Страница 11: ...2 _ _ _ _ RESET TRAINS SETTINGS BACK HOME MENU PRESETS SETTINGS BACKLIGHT SET PULSE LIFE PC BACKLIGHT FEEDBACK ON OFF SET DELAY RESET PC X RAY INFO TRAINS SETTINGS BACK SETTINGS FEEDBACK BACKLIGHT FEE...

Страница 12: ...the tube s output will decrease slowly with use If the tube is broken or the glass cracks the tube output will cease immediately 8 0 TROUBLESHOOTING SYMPTOM TEST ACTION No power on light Check batter...

Страница 13: ...he control module 3 Remove cable connecting the boards 4 Control Module Removed 1 2 3 4 Remove keyswitch connector Remove 3 screws holding processor board to top Remove processor board Disconnect ribb...

Страница 14: ...ited warranty customer must ship the entire unit or the component believed to be defective to Golden Engineering post paid Golden Engineering assumes no liability for units or components shipped until...

Страница 15: ...onds ELECTRICAL AND THERMAL CHARACTERISTICS Battery voltage 18 volts 20 volts Battery type Li Ion Battery recharge time One Hour Current draw 20 amps 18 20 volt battery Current ow 0 25 mA Temperature...

Страница 16: ......

Отзывы: