WG600S Industrial Basic Variable-Speed Peristaltic Pump
1) Turn on the power switch. The LED display will be on.
2) Make sure the
DIRECTION
switch is on middle position.
3) Press
SPEED ADJUST
dial to change the mode to Internal Control
Mode.
4) Rotate the
SPEED ADJUST
dial to adjust the speed to desired value.
5) Change
DIRECTION
switch position to desired rotating direction. The
drive starts running.
6) Press
PRIME/DISPENSE
button, the drive will run at maximum allowed
speed.
7) Change
DIRECTION
switch to the middle position to stop the drive
anytime when an accident occurs.
215
fUll
PRIME/
DISPENSE
Figure 12. Full Speed
6.4
External Control Mode (E)
On this mode, external logic level signal controls direction, start and stop,
and external analog signal controls rotation speed. The keypad is disabled.
The analog signal could be 0-5V, 0-10V or 4-20mA. By default, the signal is
0-5V. For 0-10V or 4-20mA, the jump setting on the analog signal control
board has to be changed (
0-5V (default)
0-10V
4-20mA
Figure 13. Analog Signal Control Board Setting