42
Functioning of the LCD Display
1. Indicator of the level of canal advancement
2. Visualization of the file position.
3. Indicator of the sound level: Mute, low, medium and
high.
4. Power level indicator.
5. Meter of the depth of canal advancement
6. APEX (ideal level of the root apex)
7. OVER (apex exceeding area)
2.0mm:
Beginning of the
canal easurement where
beeps will sound slowly
from 2,0mm to 1,0mm.
1.0mm:
Advancement of
the canal measurement,
where beeps will sound
continuously from 1,0mm
to 0,5 mm.
0.5mm:
Indicates
that the extremity
of the file is near
the apical region
(APEX), where
short and frequent
beeps will sound
0,5mm to 0,0mm.
0 . 0 m m :
W h e n
the apical foramen
i s r e a c h e d o r
outreached (APEX),
intermittent beeps
will sound and the
word (OVER) will light
and blink.
0.C:
Indicates that
no measurement is
being processed.
1