PRECAUTIONS, RESTRICTIONS AND WARNINGS
Sensitivity to environmental conditions predictable in normal
use situations
WARNINGS
- Endus Duo R needs special care regarding the electromagnetic compatibility and needs
to be installed and positioned to work according to information about electromagnetic
compatibility provided in this manual.
- The RF mobile communications equipment and portables can affect the electromedical
equipment.
- The use of a cable, transducer or any other accessory different from those specified
in this manual, and/or the substitution of internal components in this equipment
may result in raising the emission or reducing the electromagnetic immunity of the
equipment.
- It is not suitable that Endus Duo R be used in the proximity with or piled over another
equipment. In case there is a need of using the equipment near another one or piled,
it’s appropriate that the Endus Duo R equipment be observed to check if it’s working
normally in the configuration for which it’ll be used.
The equipment should be transported and stored with the following observations:
- With care, not to suffer from falls and from receiving impacts.
- With humidity protection, not to be exposed to rain, water drops or wet floor.
- To keep in a protected from rain and direct sun place and in its original package.
- When transporting it, do not move it through irregular surfaces and protect the package
from direct rain and respect the maximum piling informed on the external part of the package.
Environmental conditions for transportation or storage:
- Environmental temperature range for transportation or storage -29ºC to +60ºC.
- Relative humidity range for transportation or storage 20% to 90%.
- Atmospheric pressure range 500hPa to 1060 hPa (375 mmHg to 795 mmHg).
Environmental conditions of packing (between operations):
- Environmental temperature range for packing -10ºC to +55ºC.
- Recommended environmental temperature range +15ºC to +30ºC.
- Relative Humidity Range for packing - 30% to 75%.
- Atmospheric pressure range700 hPa to 1060 hPa (525 mmHg to 795 mmHg).
Operation environmental conditions:
- Working environment temperature range +10ºC to +40ºC.
- Recommended environmental temperature range +21ºC to +26ºC.
- Working relative humidity range 0% to 75%.
- Atmospheric pressure range 700 hPa to 1060 hPa (525 mmHg to 795 mmHg).
- Operating Altitude: ≤2000m.
Transportation, storage and operating conditions
ATTENTION
The equipment keeps its safety condition and efficiency, since kept (stored) as
mentioned in this operation instruction. That way, the equipment will not lose or
change its physical and dimensional characteristics.
Содержание Endus Duo R
Страница 1: ......
Страница 12: ...12 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Страница 13: ...13 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Страница 14: ...14 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Страница 15: ...15 Emissões eletromagnéticas ESPECIFICAÇÕES TÉCNICAS ...
Страница 66: ...66 ...
Страница 67: ...67 ...
Страница 68: ...68 ...