B-9
FIELD CALIBRATION
Quitting FCM
Quit And Save Changes
1) Press and hold Button Two
to save CAL data.
2) Double press Button One
to exit FCM.
Note: When all ranges have been zeroed, calibrated
correctly, CAL data saved and followed by CAL
mode exit, the new CAL DUE date will be set to
12 months from now. (This can be altered to a
different frequency, via the set-up program, e.g. 6
months from now. Contact GMI for details).
Quit Without Saving Changes
1) Double press Button One
to exit FCM.
Note: When you exit the FCM without saving the new
CAL data, the old calibration data and calibration
date remains in the instrument memory.