• Store test strip packages in a cool dry place. Keep away from direct sunlight
and heat.
• After removing a test strip from the vial, immediately close the vial cap
tightly.
• Touch the test strip with clean, dry hands.
• Use each test strip immediately after removing it from the vial.
• Write the opening date on the vial label and discard the remaining test
strips after 6 months.
• Do not use test strips beyond the expiry date. This may cause inaccurate
results.
• Do not bend, cut or alter a test strip in any way.
• Keep the strip vial away from children since the cap and the test strip may
be choking hazards. If swallowed, promptly seek medical assistance.
For further information, please refer to the GlucoCheck test strip package
insert.
Important Control Solution Information
• Use only GlucoCheck control solutions with your meter.
• Do not use the control solution beyond the expiry date or 3 months after
first opening. Write the opening date on the control solution bottle and
discard the remaining solution after 3 months.
• It is recommended that the control solution test be performed at room
temperature 20°C to 25°C. Make sure your control solution, meter-, and test
strips are at this specified temperature range before testing.
• Shake the bottle before use, discard the first drop of control solution, and
wipe off the dispenser tip to ensure a pure sample and an accurate result.
• Store the control solution tightly closed at temperatures between 2°C to
30°C. Do
NOT
freeze.