ALARMS
Certifications CE MDD Certificate
The sound and acoustic signals are in compliance with directive 60601-1-8.
Meaning
Correct power emission
The treatment proceeds
correctly
Cable not connected
Check that the cable is
connected to the right
output.
Device in pause
Press OK to continue the
treatment.
No contact
Check that gel is present on
the hand of the handpiece
No contact disabled
It indicates that the contact
control has been disabled
NOTE:
In case the window of not connected cable should appear, verify that the device cables
and the diffuser are integral and correctly used.
If the problem should persist contact the service centers.
Contact control (no contact)
It is possible to disable the contact control (symbol is shown on the upper right corner of
the display) up to the end of the ongoing treatment by pressing for more than 3 seconds
the Fn button. To enable the control again, press the "Fn" button for more than 3 seconds.
Overheating alarm (hot)
Red led lights up + "hot" + 3“ sec. blip is appearing on the display: it means that either the
head or the device are overheating. The device goes automatically in pause and the
yellow led lights up.
Problem solution for overheating head
Add gel on the head or immerse the head into water to cool it down. When the red led
turns off, it is possible to continue the treatment by pressing the OK button. Problem
solution for overheating device Check that the internal fan has automatically turned on