10
CVL
™
ClearSteth User Manual
1. To discontinue stethoscope sounds, click on the “Disconnect” button in ClearSteth
2. To disconnect from the connection,
click on the “x” button in the top right
corner of the CVL Connect window.
Use filters to focus on a
specific frequency range.
This affects local audio only,
enabling providers to change
what they hear.
The Equalizer Dialog allows the user to change the way
the audio sounds. “Enable Equalizer” selected to apply the
changes.
Amplitude Limiter:
The Amplitude Limiter allows the user to set a maximum audio level. This helps
prevent unwanted loud noises, such as those created when the stethoscope is moved, from disturbing
the user. When required, this should be set by turning the slider down (to the left) until the heartbeat
disappears, then moving it slightly back (to the right) until it reappears.
Min:
Lowest maximum volume level.
Max:
Highest maximum volume level.
EQ Type:
Applies selected equalization to enhance audio quality.
Bell:
Maintains lower frequency range for heart and bowel
sounds.
Diaphragm:
Maintains higher frequency range for breath
sounds.
Extended:
Maintains full range.
Listen locally, if not connected
to another location.
Connect button to begin
sending auscultation sounds
Mute stethoscope
Controls