CVL

™ 

ClearSteth User Manual

1. To discontinue stethoscope sounds, click on the “Disconnect” button       in ClearSteth

2. To disconnect from the connection, 

click on the “x” button in the top right 

corner of the CVL Connect window.  

Use filters to focus on a 

specific frequency range. 

This affects local audio only, 

enabling providers to change 

what they hear.

The Equalizer Dialog allows the user to change the way 

the audio sounds. “Enable Equalizer” selected to apply the 

changes.

Amplitude Limiter:

 The Amplitude Limiter allows the user to set a maximum audio level. This helps 

prevent unwanted loud noises, such as those created when the stethoscope is moved, from disturbing 

the user. When required, this should be set by turning the slider down (to the left) until the heartbeat 

disappears, then moving it slightly back (to the right) until it reappears.

Min:

 Lowest maximum volume level.

Max: 

Highest maximum volume level.

EQ Type:

 Applies selected equalization to enhance audio quality.

Bell: 

Maintains lower frequency range for heart and bowel 

sounds.

Diaphragm:

 Maintains higher frequency range for breath 

sounds.

Extended:

 Maintains full range.

Listen locally, if not connected 

to another location.

Connect button to begin

sending auscultation sounds

Mute stethoscope

Controls

Содержание CVL ClearSteth

Страница 1: ...User Manual Beta...

Страница 2: ...6020087 MAN 600380 Rev C 1 CVL ClearSteth User Manual Contents About GlobalMed 1 Specifications and Requirements 2 Installation and Configuration 2 5 Connecting Stethoscopes 6 Connect 7 9 Controls 10...

Страница 3: ...video conferencing and video streaming technologies enable multiple specialists and other caregivers to be more effective in providing patient care and more productive and efficient in managing patien...

Страница 4: ...ltation exam ClearSteth was developed with the end user in mind combining high quality audio and ease of use The ClearSteth application has the ability to function as either the Sender or Receiver sid...

Страница 5: ...otherwise it will need to be acknowledged The installer MUST run under admin level privileges If using a command line to run the CVL ClearSteth x x x exe you can use exenoui qn switches which will rem...

Страница 6: ...should be configured by qualified IT personnel only The Connectivity Tab configures the audio connection type If multiple connection types are selected ClearSteth will attempt connections in the foll...

Страница 7: ...nd 3478 for the Port b For the TCP TURN enter st globalmed com for the IP address and 3479 for the Port 4 Local IP Tab Note The connectivity tab should be configured by qualified IT personnel only The...

Страница 8: ...ful driver install will be provided via the machine s operating system 3 Start ClearSteth On the Audio tab navigate to the Audio Input Device drop down list Open the device list and select the device...

Страница 9: ...n The following is a description of the procedures both users must follow to successfully establish a CVL ClearSteth connection To configure and initiate a CVL ClearSteth connection from the sending s...

Страница 10: ...the file attachment Once opened a CVL Connect connection will be established and reflected in the drop down list of connected users as Remote User 4 Start ClearSteth by selecting the desktop icon Ensu...

Страница 11: ...sent from the sending side Open the file attachment labeled invitation cvl CVL Connect will start and display an entry for Host Patient 2 Configure ClearSteth with the appropriate Audio Output device...

Страница 12: ...The Amplitude Limiter allows the user to set a maximum audio level This helps prevent unwanted loud noises such as those created when the stethoscope is moved from disturbing the user When required t...

Страница 13: ...an end user an error message appears When an error message appears upon selection of the Connect button both users need to configure ClearSteth with an output input device Specifically the receiving s...

Страница 14: ...itiate upon start up of ClearSteth This can occur when a user is attempting to conduct both a video conference and ClearSteth simultaneously In order to concurrently stream audio from both the video c...

Страница 15: ...ry high frequencies those who can will have difficulty discerning minute subtleties that could impact diagnosis and ultimately care delivery In order to prevent soundcard deficiencies from impacting C...

Страница 16: ...select Properties 4 Navigate to the Advanced tab and locate the Default Format section Verify that the cur rently selected dropdown option reads 16 bit 48000 Hz DVD Quality If it does not open the dr...

Страница 17: ...rior of the package Unauthorized returns refused shipments and authorized returns of nondefective merchandise after the above stated return periods are subject to additional charges Limited Liability...

Страница 18: ...2002 2017 GlobalMedia Group LLC All Rights Reserved GMR16020087 MAN 600380 Rev B 16 Notes...

Страница 19: ...2002 2017 GlobalMedia Group LLC All Rights Reserved GMR16020087 MAN 600380 Rev B 17 Notes...

Страница 20: ...fax telemed GlobalMed com www GlobalMed com 2002 2017 GlobalMedia Group LLC DBA GlobalMed All Rights Reserved Additional copyright and patent information can be found at www GlobalMed com legal MAN 6...

Отзывы: