5. Configuration Parameters Setup
1. When the scale is off, press and hold ZERO/ON/OFF and PRINT/HOLD keys until
‘
C0NF
’ is shown, which indicates that the scale is in Configuration parameter setup
mode.
2. During setup mode, press the UNIT key to change the flashed digits, and use the
HOLD/PRINT key to shift the flashed position. Press the TARE key to confirm the
flashed digits. Press the ZERO/ON/OFF key to exit this mode.
3. Summary of Configuration Parameters Setting:
Para-
meter
Option
Remark
setting
C1
100 - 20000
Display Resolution for main weight unit:
100–20000
800
C2
100 - 25000
Display Resolution for second weight unit (if main unit is
kg or lb, then second unit is lb or kg): 100–25000
Note: it must be equal or less than 1.25×C1
910
C3
0, 1, 2
Division select: 0 – 1; 1 – 2; 2 – 5;
2
C4
0 - 5
Decimal point:
0 - x1; 1 - x0.1; 2 - x0.01;
3 - x0.001; 4 - x0.0001; 5 - x10;
1
C5 0,
1
Main weight unit (Calibration and setting capacity weight
units): 0 – kg; 1 – lb
1
C6
0, 1, 2
Weight Units that can be chosen by UNIT key:
0-kg; 1-lb; 2-kg/lb
2
C7
0 - 9
Power-on zero-point range:
0 - calibration zero -point ±1%FS;
1 - calibration zero –point ±2%FS;
2 - calibration zero –point ±3%FS;
3 - calibration zero –point ±4%FS;
4 - calibration zero –point ±5%FS;
5 - calibration zero –point ±10%FS;
6 - calibration zero –point ±20%FS;
7 - calibration zero –point ±50%FS;
8 - calibration zero –point ±100%FS;
9 - no limitation;
9