6.

 

Pulse Out:  This output sends a pulse to an optional recording or 

monitoring device each time the pump starts, allowing the time of the 

sample event to be recorded.  Refer to the Specifications section for 

details about the output pulse. 

7.

 

Status:  This red LED indicates the status of the sampler.  While the 

sampler is waiting to be triggered, the light remains off.  Once 

triggered, the LED will blink once every two seconds while the Delay 

or Interval timers are running and while the pump is operating.  The 

LED remains lit constantly when the sample bottle is full or if the 

bottle switch is not plugged in. 

8.

 

Power:  This turns the power to the sampler on and off.  Turning the 

switch off for one second resets all functions and timers. 

9.

 

Charger:  Connect the battery charger here. 

10.

 

Water Sensor:  When an optional water sensor is plugged into this 

jack, sampling will not begin until the presence of water is detected 

(the Rain Sensor must also detect water).  If this sensor input is not 

being used, plug one of the two shorting plugs supplied with the 

sampler into this jack.  This input has no affect when triggering the 

sampler using the Trigger Input in External Trigger mode. 

11.

 

Rain Sensor:  When an optional water sensor is plugged into this 

jack, sampling will not begin until the presence of water is detected, 

usually in a rain gauge (the Water Sensor must also detect water).  If 

this sensor input is not being used, plug one of the two shorting 

plugs supplied with the sampler into this jack.  This input has no 

affect when triggering the sampler using the Trigger Input in External 

Trigger mode.  This input can also be used to take individual 

composite samples and is described in detail later in the section 

titled “External Trigger Mode”. 

12.

 

Bottle Switch:  The sample bottle float switch plugs into the control 

panel here.  For sampling to occur a float switch must be plugged in 

and the bottle must not be full.  If the Status LED remains lit all the 

time, check this connection. 

13.

 

Test Pump:  Pressing this button tests the operation of the pump.  

This button will run the pump regardless of whether the bottle switch 

is plugged in or not.  The pump runs forward for as long as the 

button is held down, then reverses for an equal amount of time or 15 

seconds, which ever is less. 

 

Содержание WS700R

Страница 1: ...700R Refrigerated Wastewater Sampler Global Water Instrumentation Inc 151 Graham Road P O Box 9010 College Station TX 77842 9010 T 800 876 1172 Int l 979 690 5560 Fax 979 690 0440 E mail globalw globa...

Страница 2: ...he Refrigerator 8 VII Composite Sampling 8 VIII Discrete Sampling 9 IX External Trigger mode 9 X Specifications 11 XI Maintenance 12 XII Troubleshooting 13 XIII Warranty 14 XIV Appendix A Sample Size...

Страница 3: ...help I WS700R Checklist a WS700R Wastewater Sampler Controller b 12V Battery Inside sampler housing c Battery Charger d Two Shorting Plugs e WS700R Manual f Refrigerator g Sample Bottle inside refrige...

Страница 4: ...ave them turn on at the same time The delay can also be used to hold off sampling to let the water source flush out debris or other contaminants After each sample the pump will reverse for 15 seconds...

Страница 5: ...come contaminated Sampler a The control panel should be placed upright b To secure the sample bottle 1 Screw the bottle caps float switch onto the sample bottle 2 Place bottle into the refrigerator 3...

Страница 6: ...ple Size Sets the size for a single composite sample The sample size settings are approximate and apply to a head height of four feet The switch position marked Full causes the pump to work as a discr...

Страница 7: ...horting plugs supplied with the sampler into this jack This input has no affect when triggering the sampler using the Trigger Input in External Trigger mode 11 Rain Sensor When an optional water senso...

Страница 8: ...ller samples put into the same bottle thus showing an average sample over time Set the Interval control to the desired time between samples do not select the External Trigger position Set the Size con...

Страница 9: ...rnally by either the Trigger In jack or the Rain Sensor input This mode is useful in applications where sampling occurs based on readings from a process controller SCADA system flow monitor auto dumpi...

Страница 10: ...y timer begins running When the delay timer has expired the pump takes a sample based on the Sample Size setting Holding the Trigger Input active past the end of the sample will cause the pump to cont...

Страница 11: ...l Cell Battery Life 1 hour continuous pumping under load Standby 3 months while still retaining enough power to run the pump to capacity Start Delay 16 time settings from 0 to 12 hours Composite Inter...

Страница 12: ...water Rinse and dry g Defrost whenever frost on the wall of the freezer compartment becomes thick Never use a sharp or metallic instrument to remove frost as it may damage the cooling coils A puncture...

Страница 13: ...ging one of the shorting plugs supplied with the sampler into the jack ii The LED is blinking Check the Start Delay and Interval settings The sampler may be waiting for the Delay timer to run out or i...

Страница 14: ...ntation Inc warrants that its products are free from defects in material and workmanship under normal use and service for a period of one year from date of shipment from factory Global Water s obligat...

Страница 15: ...raph to see how sample size changes with head height The Sample Size controls on the control panel are calibrated at a height of 4 feet Increasing the head height decreases the sample size by a predic...

Страница 16: ...veral factors such as charge condition and temperature it can generally be assumed that the sampler can deployed in the field for more than 3 months waiting to be triggered while still retaining enoug...

Страница 17: ...umber Unit 12V 5Ah Battery 00 010 Each Battery Charger 00 017 Each 2 Gal Plastic Sample Bottle 00 738 Each Bottle Cap Float Switch CD0300 Each Pickup Hose 00 546 Feet Pickup Strainer CA0300 Each Pump...

Отзывы:

Похожие инструкции для WS700R