Using Global Logger Software 

Pulse Channel Calibration 

To change pulse channel settings, refer to the 

Pulse Input

 fields in the 

Global Logger Setup

 window.  The 

pulse channel records a predetermined value each time the sensor detects a pulse signal.  This feature is of-
ten used to track rainfall levels. 

To change the displayed unit type (i.e. meters, feet, etc.), tap the 

Sensor Name or EU

 (engineering units) 

field and enter the unit type you are calibrating to.  Note that the 

Sensor Name or EU

 field only serves as a 

text label. 

To set the recorded value for a single pulse reading, tap the 

Value for Single Pulse

 field and enter the de-

sired predetermined value. 

To set the number of decimal place values to display, tap the 

Decimal Places

 field and enter the number of 

decimal places (0 – 6), or click the up/down arrows to the right of the text field to increase or decrease the 
value by 1.  After changing settings, click the 

OK

 button.  To cancel changes, click 

Cancel

Verify that 

High EU

 and 

Low EU

 fields reflect desired unit readings.  Also, verify that the 

High Raw

 field 

reflects a range of 60,000 – 65,000 and that the 

Low Raw

 field reflects a range of 10,000 – 15,000. 

Adjust the 

Decimal Places

 field according to the accuracy percentage rating.  The number of decimal 

places should not exceed the accuracy of the sensor.  After calibrating, click the 

OK

 button.  To cancel 

changes, click 

Cancel

Содержание GL400

Страница 1: ...Page 1 Global Water 800 876 1172 www globalw com 1 05 2005 Copyright Global Water Instrumentation Inc 2002 GL400 Datalogger User s Guide...

Страница 2: ...is product manual please be aware of the following terminology and formatting standards Bolded terms i e bold indicate either a keyboard key a word or icon that appears on your com puter monitor scree...

Страница 3: ...irst and last items in the string separated by For example the phrase Click on File New Folder means click on File select New and click on Folder When the documentation directs you to exit a program o...

Страница 4: ...inuously Real Time Values 20 Get Settings 20 Get History Data 21 Clear Memory 23 Synchronize Time 23 Setup 23 Datalogger Name and Sensor Warmup Time 24 Analog Channel Calibration 25 Pulse Channel Cali...

Страница 5: ...ain Gauges Contact us for unique special and custom modifications to meet the requirements of your application The GL400 Datalogger is designed to accept any 4 20 mA sensors Twisted pair 2 wire and 3...

Страница 6: ...checklist to ensure that you have received everything needed to operate the GL400 Note there are two different varieties of dataloggers the GL400 7 1 and GL400 1 1 Use the checklist appropriate to you...

Страница 7: ...for GL400 1 1 GL400 1 1 2 Channel Datalogger User Manual Communication Cable RS232C COM Cable Global Logger Software 3 1 2 Floppy Diskette Communication Cable Global Logger Software User Manual Datalo...

Страница 8: ...he GL400 The COM port can connect to a PC se rial port using the included Communica tion Cable COM port Access the included 12V 2 AH lead acid gell cell directly from inside enclosure Chapter 1 The GL...

Страница 9: ...fe male connector to your PC and the other to your Datalogger GL400 7 1 GL400 1 1 Introduction Connect sensors to the GL400 by threading sensor cables through the enclosure s strain reliefs 8 on the G...

Страница 10: ...ify your system time in Windows click Start Set tings Control Panel from the task bar Double click the Date Time icon Before you install the software check the accuracy of the time and date on your co...

Страница 11: ...n Exit the control panel Double click the My Computer icon on your desktop Double click the floppy icon normally listed as 3 floppy A Insert the 3 1 2 floppy diskette labeled Global Logger Software Ve...

Страница 12: ...Page 12 Global Water 800 876 1172 www globalw com Click on the file Global Logger exe Click Edit Copy from the menu bar Chapter 2...

Страница 13: ...To save to the directory Global Logger located on your main C hard drive although C is normally the main system drive the designation can vary from system to system start by clicking the down carrot...

Страница 14: ...obal Water 800 876 1172 www globalw com Double click on the new Global Logger directory Click Edit Paste from the menu bar A Copying dialog box should now display Wait for the copy process to complete...

Страница 15: ...ver if the datalogger is submerged in water it will corrode the electronics and cause it to stop working properly Please ensure that the datalogger will be clear from flooding and out of direct rain f...

Страница 16: ...source to the Ext terminals located be low the internal battery Batt terminals battery terminals analog channels 1 7 pulse channel Chapter 2 To connect a sensor thread sensor cable through an enclosu...

Страница 17: ...nnected to the logger and is accessible through the Global Logger software pack age See Global Logger Startup page 18 Copyright Global Water Instrumentation Inc 2001 For pulse sensors polarity is not...

Страница 18: ...hich your Global Logger exe file is located and double click on the file If you have not installed the Global Logger software or if you want to create a desktop icon see Global Logger Software Install...

Страница 19: ...aphs indicate zero to full scale range of sensor for the eight analog channels of the datalogger For example the channel 4 graph in the following screen shot displays a 120 foot sensor with a reading...

Страница 20: ...Copyright Global Water Instrumentation Inc 2001 Global Logger Functions Sample Continuously Real Time Values To display real time values each second select the Sample Continuously checkbox Readings wi...

Страница 21: ...ta point for every channel by Date Time and reading value Channel 1 readings are displayed in the column immediately to the right of the Date Time column and additional channels consecutively to the r...

Страница 22: ...eld and type the desired name However do not change the csv extension This extension ensures the file will be recognized as a comma sepa rated value CSV file You can open CSV files in spreadsheet prog...

Страница 23: ...ynchronize your GL400 s internal clock to your computer s clock click the Synchronize Time but A Standby graphic will display in the middle of your monitor screen while synchronizing Setup Note that y...

Страница 24: ...ors before the datalogger takes a reading For example you set Global Logger to a sensor warm up time of 3 seconds and to record readings every 20 minutes at fixed intervals If the process begins at 12...

Страница 25: ...r the maximum unit range of the sensor 120 for a 120 level sensor in this example in the highest value text field and click the Next button Change the Sensor Type or EU engineering units field to the...

Страница 26: ...and lab calibrations wait at least five seconds for the sensor to adjust to 0 sensing levels Click the Finish button See Appendix B Worksheets pages 36 Note that if performing an envi ronmental calibr...

Страница 27: ...lse reading tap the Value for Single Pulse field and enter the de sired predetermined value To set the number of decimal place values to display tap the Decimal Places field and enter the number of de...

Страница 28: ...rrows Set the recording interval time unit by clicking the down carrot of the time unit field not labeled and clicking Seconds Minutes Hours or Days Chapter 3 Note that if you choose to record at fixe...

Страница 29: ...t the bottom of the Global Logger Setup menu The bar should change to reflect the latest exception setting The setting in these screenshot examples sets the logger to record a single reading after rea...

Страница 30: ...om Chapter 3 After calibrating and or changing interval recording settings click the OK button To cancel changes click Cancel You may also manually change GL400 settings directly using a terminal prog...

Страница 31: ...ger will not work reliably if the battery voltage is below 10V If the battery is allowed to drain below 9V it may not charge back to 12V when the charger is applied The battery can be recharged using...

Страница 32: ...orm the following steps Verify that the communication cable is securely connected to the data logger and to your computer Check the battery voltage on the logger Make sure that your computer is correc...

Страница 33: ...Email globalw globalw com When calling for tech support please have the following information ready 1 Model number 2 Unit serial number 3 P O number the equipment was purchased on 4 Our sales number...

Страница 34: ...urs or 1 to 256 days Time Accuracy 0 0015 Time Format Month Day Year Hour Minute Second Communication Speed 38 400 kbs Data Transfer 2MB 4 min Type RS232C Serial Electrical Battery 12V Sealed Recharge...

Страница 35: ...GL400 Detailed Specifications Physical Properties Severe Environment Enclosure Loaded Length 254mm 10 Width 191mm 7 5 Height 114mm 4 5 Weight 2 268 kg 5lbs Vent Port Open or Closed Lockable Entry Port...

Страница 36: ...new limits on the bar graph First take the two coordinates that were found during calibration High EU __________ High Raw __________ and Low EU __________ Low Raw __________ Now calculate the slope u...

Страница 37: ...are not exact You may need to measure for precise distances Total Depth _____________ Next find the maximum Sensor Depth range of the sensor In the Calibration section of the software you can find thi...

Страница 38: ...nd Description D Set Date T Set Time Get number of data points h Get first historical data point a Get historical data point again X Abort i Get next historical data point increment E Erase r Get real...

Страница 39: ...Recharging the Battery Disconnect the datalogger s red and blue spade lugs from the battery Ensure that the battery charger is not plugged in Disconnect spade lugs from battery Connect the charger s...

Страница 40: ...g the charger and check the battery voltage using a multimeter If the battery voltage has increased but not reached 12V it may need to be left on the charger longer If the battery voltage has not chan...

Страница 41: ...e protected this allows the battery to remain charged while the external source has power CAUTION The enclosure must be vented or gasses could build up inside the case causing a rupture Alternatively...

Отзывы: