background image

Global Specialties 

Calibration 

Humidity 

1. Turn the unit off.  Pres and hold 

MAX MIN, HOLD 

and 

•c 

buttons. While these

buttons are pressed, turn the power on.  Release the buttons  and all the segments

on the LCD will blink.

2. Within three seconds, press the 

'F 

and 

t.REL 

buttons at the same time to enter

the calibration mode.

3.  In calibration mode, the humidity reading will blink and "CAL 1" will appear on the 

second display. 

4. Insert the humidity probe into the standard humidity cavity of 32.8% RH at 25 

·c.

Let the system stabalize for 20 minutes then press the 

MAX MIN 

button to create

the calibration data.  If the value is out of tolerance, the meter will beep twice and

remain in "CAL 1" mode.  If the value is within tolerance the meter will go directly

into "CAL2" mode and display "CAL2" on the second display

5. Insert the humidity probe into the standard humidity cavity of 75.3% RD at 25 

·c

and let the system again stabalize for 20 minutes.  Press the 

MAX MIN 

button

to create the calibration data.  The meter will beep twice if the

• 

value is out of

tolerance.  The meter will exit cal bration mode if the value is within tolerance.

T1  Temperature 

1. Turn the unit off.  Pres and hold 

MAX MIN, HOLD 

and 

•c 

buttons. While these

buttons are pressed, turn the power on.  Release the buttons and all the segments

on the LCD will blink.

2. Within three seconds, press the 

'F 

and 

HOLD 

buttons at the same time to enter

the calibration mode.

3. In calibration mode, the temperature reading will blink and "CAL 1" will appear on

the second display.

4. Insert the probe into the standard chamber of 0 

·c 

(32 'F).  Let the system

stabalize for 20 minutes then press the 

MAX MIN 

button to create the calibration

data.  If the value is out of tolerance, the meter will beep twice and remain in

"CAL 1" mode.  If the value is within tolerance the meter will go directly into "CAL2"

mode and display "CAL2" on the second display

5. Insert the humidity probe into the standard chamber of 40 

·c 

(104' F) and let the

system again stabalize for 20 minutes.  Press the 

MAX MIN 

button to create the

calibration data.  The meter will beep twice if the value is out of tolerance.  The

meter will exit calibration mode if the value is within tolerance.

Содержание GNV-720

Страница 1: ...GNV 720 GNV 725 Humidity Temperature Meters User Manual Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...

Страница 2: ...ion Using the Software Calibration Instructions Warranty Service Information 2019 Global Specialties All rights reserved 2 2 3 3 4 5 6 7 7 8 9 10 11 Unauthorized duplication of Global Specialties docu...

Страница 3: ...rged at any time Dry thoroughly before attempting to make measurements Do not use solvents or expose to solvent fumes as they may cause deterioration or damage Do not operate In the presence of noxiou...

Страница 4: ...facilities in the disposition of this product Caution Refer to the operating instructions Introduction The GNV 720 and GNV 725 are digital humidly temperature meters that measure relative humidity and...

Страница 5: ...1 C CT32 40 J 200 Fto 200 F 1 F 200 F to 2498 F 1 F Accuracy Humidity 2 5 RH at 25 C Temperature T1 e0 7 C e1 4 FJ Temperature T2 0 5 reading 1 C 0_5 reading 2 F Response Time Humidity 75s Temperatur...

Страница 6: ...output connector RS 232 13 AC power adapter connector 14 Tripod mounting hole 15 Battery cover E il 0 Cl AREL MAX _ K BBB B ffi d n ru111111iI T2 8BfJ 8 88 8 Bi c F Celsius and Fahrenheit indicators R...

Страница 7: ...both be displayed while MAX and MIN are both blinking To exit this mode simply hold the MAX MIN button for two seconds To view the change in values press the c REL button This uses the current value a...

Страница 8: ...save settings Pressing the Power button at any time will cancel the process To display the time press the TIME button The time will be displayed with the year at the top the month and day at the botto...

Страница 9: ...by typing in the sample rate box GNV 725 Data logging can be selected by clicking the appropriate menu at the top of the tool bar Data will automatically be downloaded Once the data is loaded succesf...

Страница 10: ...The meter will beep twice if the value is out of tolerance The meter will exit cal bration mode if the value is within tolerance T1 Temperature 1 Turn the unit off Pres and hold MAX MIN HOLD and c but...

Страница 11: ...tory conditions press the and hold the power button and HOLD buttons together It is recommended to first insert the probe into the standards for 20 minutes before initiating the calibration procedure...

Страница 12: ...451001 000 Rev A Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...

Отзывы: