Global Focus SILENCER S12V Series Скачать руководство пользователя страница 10

Using The S12V 

 

S12V Service Manual 

Page 3-2 

There are four LEDs to indicate the selected speed. A blinking LED is an indication that 
an error has occurred. 

MINUTES 

 

There are four LEDs to indicate the selected time in minutes. A blinking LED normally 
indicates that the unit has reached set speed. If the RPM / RCF LED is also blinking an 
error has occurred. 

 

     

WARNING 

Do not operate this equipment in a combustible 
atmosphere! 

 

O

PERATION

 

 

Chapter 1 must be read and the S12V must be setup accordingly before continuing.   

The set speed and time defaults to the last used speed and time setting. The settings can be 
changed using the SPEED and TIME buttons. 

Open the lid and load your samples, taking care to maintain a balanced load. 

Close the lid and press the START button. The lidlock will engage, and the centrifuge will 
accelerate to the set speed. Upon reaching the set speed the time LED will start blinking. 

When the set time has elapsed the time LED will stop blinking and the centrifuge will 
decelerate to a stop. 

When the rotor has come to a complete stop the lidlock will release. 

 

O

PERATION 

T

IPS

 

 

   

Do not drop or slam the lid! 

   Use the emergency lid lock release only for emergencies. 

   Only use the supplied emergency lid lock release tool. 

   Unplug the centrifuge before using the emergency lid lock release tool. 

   Leave the lid open between cycles. 

 
If there are any questions regarding operation please contact GFMD, Ltd. at 800-323-4306. 
 
 

Содержание SILENCER S12V Series

Страница 1: ...S12V Series Service Manual Service Manual Proudly designed and manufactured in the great United States...

Страница 2: ...rights to reproduce or disclose its contents in any form Global Focus Marketing Distribution Ltd herein known as GFMD Ltd retains all rights to this manual and its contents Reproducing and or disclosu...

Страница 3: ...apter 5 Troubleshooting 5 1 Status Codes 5 1 Error Reset 5 1 Troubleshooting Table 5 1 Chapter 6 Functional Description 6 1 AC Power Distribution 6 1 DC Power Distribution 6 1 Operator Interface Circu...

Страница 4: ...idlock Assembly 8 3 Chapter 9 Warranty 9 1 Appendix A Specifications A 1 Performance Specifications A 1 Unit Specifications A 2 Appendix B Service Log A 3 Appendix C Replacement Parts List A 4 Appendi...

Страница 5: ...e 3 1 Control Panel Layout 3 1 Figure 7 1 Wiring Diagram 7 1 Figure 7 2 Display Control PCB Schematic 7 2 Figure 7 3 Display Control PCB Layout 7 3 Figure 8 1 Front Panel Retaining Screws 8 2 Figure 8...

Страница 6: ...t damage to the centrifuge be sure to read this manual before continuing with setup or operation of this centrifuge For smooth operation and long equipment life choose a stable level bench top Avoid s...

Страница 7: ...amage to electronic circuits CAUTION Operation of centrifuge on a power strip increases ground resistance and can cause damage to electronic circuits ROTOR INSTALLATION Insert the rotor onto the motor...

Страница 8: ...This design also promotes long bearing life and eliminates brush maintenance thereby reducing operating costs Excellent stability against imbalance is achieved due to special vibration absorbing drive...

Страница 9: ...power off does not remove power to the centrifuge Remove all power to the unit by unplugging the power cord CONTROLS START If pressed when the unit is in the STANDBY mode the START button will start t...

Страница 10: ...using the SPEED and TIME buttons Open the lid and load your samples taking care to maintain a balanced load Close the lid and press the START button The lidlock will engage and the centrifuge will ac...

Страница 11: ...rotor chamber to rise to the bottom of the motor seal may result in serious damage to the centrifuge The rotor must be removed from the centrifuge for disinfecting Clean tube breakages and spills imm...

Страница 12: ...1 3 Do not attempt to open the lid until the unit has stopped ERROR 2 1 Indicates no power is getting to the motor ERROR 2 2 Indicative of a momentary drop or loss of line voltage ERROR 2 3 Indicative...

Страница 13: ...or misaligned Check the lid lock safety interlock switch for proper alignment and continuity SYMPTOM Will not start when the START button is pressed and ERROR 2 1 is displayed POSSIBLE CAUSES POSSIBLE...

Страница 14: ...complete stop see suggestions for the centrifuge not starting and giving an ERROR 1 1 error If someone attempted to open the lid during the run cycle please wait until the unit has come to a complete...

Страница 15: ...nnection is secure the unit is not plugged into a power strip and the power circuit is not overloaded SYMPTOM The sample temperature is exceeding 39 C POSSIBLE CAUSES POSSIBLE SOLUTIONS The rotor cham...

Страница 16: ...ply and the vent fan DC POWER DISTRIBUTION The output of the 24VDC Power Supply is applied to the Display Control PCB The positive output goes to PLUS1 VIN and negative output goes to NEG1 VSS The com...

Страница 17: ...applied to the gate of Q1 Q1 applies VSS to the solenoid After a time delay set by C7 and R28 the unlocking voltage turns off The solenoid holding voltage is supplied by Q2 Q2 applies VSS to the solen...

Страница 18: ...eracts the PWM pulse to limit current draw of the motor This voltage is also applied to ADC pin 22 of U1 to supply current draw feedback This feedback is used to control the PWM ramp during accelerati...

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ...the power cord EQUIPMENT The following equipment is useful or necessary in the repair of the S12V centrifuge A Test Equipment 1 Tachometer 2 Multimeter 3 Ammeter capacity to 10 amps B Tools 1 Screwdri...

Страница 23: ...tor chamber see Figure 8 2 and then pull it from around the motor Unplug the red green and black motor power wires Remove the four 5 16 hex motor securing nuts from the bottom of the motor mount plate...

Страница 24: ...nting screws Remove the Display Control P C B Install replacement Display Control P C B by reversing the removal procedure LIDLOCK ASSEMBLY Unplug the power cord Remove the four Phillips front panel r...

Страница 25: ...cuits other than those specified on the plate affixed to the instrument insufficient or excessive electrical supply operation on emergency or backup generators abnormal mechanical or environmental con...

Страница 26: ...500 RPM Maximum RCF S12V HZ w 8 x 10ml Horizontal Rotor 2 418 x g S12V FA w 10 x 10ml FA Rotor 2 802 x g Approximate acceleration rate 0 Max RPM 15 seconds Approximate deceleration rate Max 0 RPM S12V...

Страница 27: ...x 16 5 d x 11 5 h Spill resistant control panel Completely enclosed rotor chamber High strength alloy rotor chamber See through high strength lid Dual safety interlock to ensure lid is closed and lock...

Страница 28: ...S12V Service Manual Page A 3 Appendix B Service Log Service Log Date Service Performed By...

Страница 29: ...S12V 2 ea SIL00512 Latch Lidlock 1 ea SIL00642B Lid S12V 1 ea SIL00661 Lidlock Assembly 1 ea SIL00672 Switch Lid Interlock part of SIL00672 1 ea SIL02456 Switch Lock Interlock part of SIL00672 1 ea S...

Страница 30: ...d for your unit or the lid safety interlock switch may become damaged If the interlock switch is damaged the unit can t detect the lid being closed and therefore won t start To use the release tool 1...