Control mode operation

 

Manual operation provides direct (time proportioned % time) control from -100% to 100%.  A negative value is 
allowed only when Ot 1=Cool.  Automatic operation provides sensory feedback ON/OFF or PID control.  When 
the operation transfers from automatic to manual operation, the power level from automatic operation is retained 
and restored to the previous set point. 
 
The % LED indicates manual operation.  The LED is on when in Manual operation and off when in AUTO 
operation.  When the LED is flashing, press the 

Infinity

 key again within 5 seconds to complete the change in 

operation.  It the sensor is open and LOC=0, 1 or 2, the control switches to Manual operation if the output was 
stable before the break occurred. 
 

Tuning-Automatic 

 
It is recommended that auto-tuning be used each time a new process is initiated. If the process is being 
duplicated, the parameters learned in the initial auto tune setup are stored in the memory of the control and the 
auto-

tune procedure isn’t required in most applications. 

 
Auto-tuning: This control can automatically tune the PID parameters to fit the characteristics of your particular 
thermal system.  
 
Auto-tune can only be used when Output 1 setting is hEAt and heat algorithm (h.Ag, page 5) is set for PID mode. 
The heat algorithm setting is set for PID from the factory. When the auto-tune sequence has started, the lower 
display will flash between tun1-attn and the main operation screen. 
 
Changing the set point, while in auto-tune mode, will not change the set point value. The displayed value will 
change but the control has locked in the original set point value tell the auto tune in complete or has been turned 
off by the user. Once the control has finished "learning" the system, it returns to standard PID control with the 
values automatically set. Auto-tuning is complete within 80 minutes.  In order for the control to successfully 
complete auto-tune, the process must cross the 90% set point value four times within 80 minutes after auto-tune 
has started. If this does not happen within the 80 minute time limit, Pb1 remains at 0 and the control will operate 
with the factory default PID values. 
 

 

 
 
 
 
 
 
 
 

Содержание 104A PLSM112

Страница 1: ...DIGITAL TEMPERATURE CONTROLS PLSM SERIES Models 104A PLSM112 104A PLSM112K 104A PLSM112T15A 120V 50 60 Hz 104A PLSM212 104A PLSM212K 104A PLSM212T15A 120V 50 60 Hz 104A PLSM312 104A PLSM312K 104A PLS...

Страница 2: ...tch to the ON position Wait five seconds for the control unit to energize The control is shipped from the factory with the display reading in degrees C If a display in degrees F is desirable press the...

Страница 3: ...Reset 0 00 to 99 99 repeats per minute Integral 0 00 to 99 99 minutes per repeat Rate or Derivative 0 00 to 9 99 minutes Operator Interface Advance Infinity Up and Down keys and ON OFF switch Dual fo...

Страница 4: ...d Default 25 0 F or 14 0 C C Pb Cool Proportional Band Default 25 0 F or 14 0 C ti Time Integral Default 180 td Time Derivative Default 0 o tb1 Time Base Output 1 Default 1 0 or 20 0 o tb2 Time Base O...

Страница 5: ...FF h Ag Heat Algorithm Default PID hSC Hysteresis Heat Cool Default 3 0 F or 2 0 C C Ag Cool Algorithm Default oFF A ty Alarm Type Default oFF A hy Alarm Hysteresis Default 1 0 A LA Alarm Latching Def...

Страница 6: ...oftware Build Read Only Sn Serial Number Read Only dAtE Date of Manufacture Read Only USr r User Restore Settings nonE 2onE Zone Default oFF Chan Channel Default oFF CAL FCty Calibration Menu Operatio...

Страница 7: ...e procedure isn t required in most applications Auto tuning This control can automatically tune the PID parameters to fit the characteristics of your particular thermal system Auto tune can only be us...

Страница 8: ...instructions on how to access and navigate through the setup menu If excessive temperature overshoot is occurring adjust the value to 50 and start the heating process from ambient room temperature If...

Страница 9: ...a set point by turning the output full On below set point and full Off above set point Output Action in response to difference between set point and process variable Overshoot Condition where temperat...

Страница 10: ...apparatus due to improper installation or through attempts to operate the apparatus beyond its rated capacity intentional or otherwise Limitations of Warranties APART FROM SUCH WRITTEN STATEMENT OF W...

Отзывы: