OPERATING AND SAFETY INSTRUCTIONS 

for 

DIGITAL TEMPERATURE CONTROLS (PLSM SERIES) 

 

 

Models:  

104A PLSM112; 104A PLSM112K; 104A PLSM112T 15A; 120V, 50/60 Hz. 

  

104A PLSM212; 104A PLSM212K; 104A PLSM212T 15A; 120V, 50/60 Hz. 

  

104A PLSM312; 104A PLSM312K; 104A PLSM312T 15A; 120V, 50/60 Hz. 

  

104A PLSM124; 104A PLSM124K; 104A PLSM124T 10A; 240V, 50/60 Hz. 

  

104A PLSM224; 104A PLSM224K; 104A PLSM224T 10A; 240V, 50/60 Hz. 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание 104A PLSM112

Страница 1: ...DIGITAL TEMPERATURE CONTROLS PLSM SERIES Models 104A PLSM112 104A PLSM112K 104A PLSM112T15A 120V 50 60 Hz 104A PLSM212 104A PLSM212K 104A PLSM212T15A 120V 50 60 Hz 104A PLSM312 104A PLSM312K 104A PLS...

Страница 2: ...tch to the ON position Wait five seconds for the control unit to energize The control is shipped from the factory with the display reading in degrees C If a display in degrees F is desirable press the...

Страница 3: ...Reset 0 00 to 99 99 repeats per minute Integral 0 00 to 99 99 minutes per repeat Rate or Derivative 0 00 to 9 99 minutes Operator Interface Advance Infinity Up and Down keys and ON OFF switch Dual fo...

Страница 4: ...d Default 25 0 F or 14 0 C C Pb Cool Proportional Band Default 25 0 F or 14 0 C ti Time Integral Default 180 td Time Derivative Default 0 o tb1 Time Base Output 1 Default 1 0 or 20 0 o tb2 Time Base O...

Страница 5: ...FF h Ag Heat Algorithm Default PID hSC Hysteresis Heat Cool Default 3 0 F or 2 0 C C Ag Cool Algorithm Default oFF A ty Alarm Type Default oFF A hy Alarm Hysteresis Default 1 0 A LA Alarm Latching Def...

Страница 6: ...oftware Build Read Only Sn Serial Number Read Only dAtE Date of Manufacture Read Only USr r User Restore Settings nonE 2onE Zone Default oFF Chan Channel Default oFF CAL FCty Calibration Menu Operatio...

Страница 7: ...e procedure isn t required in most applications Auto tuning This control can automatically tune the PID parameters to fit the characteristics of your particular thermal system Auto tune can only be us...

Страница 8: ...instructions on how to access and navigate through the setup menu If excessive temperature overshoot is occurring adjust the value to 50 and start the heating process from ambient room temperature If...

Страница 9: ...a set point by turning the output full On below set point and full Off above set point Output Action in response to difference between set point and process variable Overshoot Condition where temperat...

Страница 10: ...apparatus due to improper installation or through attempts to operate the apparatus beyond its rated capacity intentional or otherwise Limitations of Warranties APART FROM SUCH WRITTEN STATEMENT OF W...