INHALATION EXPOSURE SYSTEM 

 

MODEL: 

099C A5212 (120 Volts) 
099C A5224 (240 Volts) 

 

 

 

 

Glas-Col, LLC 

711 Hulman Street, P.O. Box 2128, Terre Haute, IN 47802-0128 

Phone. (812) 235-6167, Fax. (812) 234-6975 

E-mail: 

[email protected]

 

Web:  

www.glascol.com

 

 

Содержание 099C A5212

Страница 1: ...47802 812 235 6167 FAX 812 234 6975 INHALATION EXPOSURE SYSTEM MODEL 099C A5212 120 Volts 099C A5224 240 Volts Glas Col LLC 711 Hulman Street P O Box 2128 Terre Haute IN 47802 0128 Phone 812 235 6167 Fax 812 234 6975 E mail pinnacle glascol com Web www glascol com ...

Страница 2: ...RE UNIT 4 EXTERNAL CONNECTIONS 5 SPECIFICATIONS FOR EACH COMPONENT 5 PRESENT OPERATION PROCEDURE 8 PRESENT OPERATION PROCEDURE 8 SAFETY 15 SPARE REPLACEMENT PARTS LIST 15 Limited Warranty 15 NOTE This symbol draws attention to notes It indicates important additional information and special tips on the immediate subject matter ...

Страница 3: ...d Air under pressure atomizes the suspension which is mixed with filtered room air drawn into the system by a suction beyond the decontaminate system By varying the number of micro organisms in the suspension the proportion of bacteria bearing nuclei can be varied The system permits control of the volume of air moving through the chamber per unit of time The animals are placed in a compartment mes...

Страница 4: ...ully open remove the nut and plastic strip on the threaded mounting studs that protrude on one end of each aluminum reflector Insert the pins on the U V lamp into the socket that is mounted to the opposite end of the reflector Slip the U shaped ultraviolet lamp over the stud and then the nut Tighten gently so that the plastic strip holds the lamp firmly but without excessive pressure Use caution n...

Страница 5: ...ACH COMPONENT LID A 5 8 thick acrylic plastic lid impervious to ultraviolet rays covers the chamber Three latches provide sufficient pressure on the gasket to seal the lid to the tank To open pull out on each latch To close press firmly on the red handle until it snaps ULTRAVIOLET LAMPS One U shaped ultraviolet lamp with an intensity of 4 watts at 254 Nanometers has a rated life of 1000 hours ...

Страница 6: ...ropylene tank is provided for the infection chamber The tank is airtight and all connections have been welded or securely gasketed BAFFLE A baffle plate is provided on the bottom of the basket requiring the basket to be inserted with the baffle plate toward the bottom and centered CABINET The system is housed in a stainless steel cabinet 30 wide by 24 deep by 21 tall Flowmeters controls and nebuli...

Страница 7: ... unit These filters use a fiberglass filter media that is 99 97 efficient on particles as small as 0 2 microns Two filters are used to filter incoming room air The output filter removes particles from the chamber air prior to incineration PROGRAMMABLE CONTROL A Micro logic control is used to time and control each cycle of the IES The user interface is located on the front panel and will display a ...

Страница 8: ...the tank of up to 50 SCFM The amount of vacuum is varied by adjusting the vacuum control valve INCINERATOR The incinerator is used to decontaminate the air as it leaves the system This is a single stage incinerator with internal temperature set at 600 degrees Fahrenheit PRESENT OPERATION PROCEDURE 1 Close lid Press firmly on each latch until closed 2 The following pages show the screens on the dis...

Страница 9: ...9 Rev 0 9 11 4 Press F1 to begin and the following will screen appear Screen 1 Screen 2 ...

Страница 10: ... 15 minute preset time is the recommend minimum time for the step The default times can be changed on any screen by pressing F2 and the following pop up box will appear on the screen Use the keypad to the right to enter a new time in minutes and then press the return key Return Key Screen 3 ...

Страница 11: ...11 Rev 0 9 11 5 The Nebulization entry is the next screen 6 Cloud Decay Entry Screen 4 Screen 5 ...

Страница 12: ...V lights are ON By pressing F3 the UV lights will be off during this step You can press the F3 at anytime during this step to turn the UV lights on or off 8 After all entries are done the following screen will appear Press F1 to start the run Screen 6 Screen 7 ...

Страница 13: ...irflow over the incinerators heating elements Cooling down the elements will greatly extend the life of the incinerator and this step cannot be changed 10 After the Cool Down step the process is now complete the alarm will sound and the following messa Screen 8 Screen 9 ...

Страница 14: ... as if the vacuum control must be increased to maintain the rated flows The filter is accessible from the rear of the unit and is labeled on its top surface Disconnect power before removing the rear panel The filters are immediately visible when the back is removed and are the smaller capsules OUTPUT HEPA FILTER Located in the middle of the unit and is labeled on its top surface Again this filter ...

Страница 15: ...om defects in material and workmanship for one year and agrees to repair or replace without charge any products found defective upon examination at the factory With proper care and operation Glas Col products will give long and efficient service Chemical spillage overloading and general misuse will greatly reduce the service life Glas Col is not responsible for damage to apparatus due to improper ...