Functions 

 
Disinfection 

 

Flowmeter 

The  flowmeter  is  used  to  adjust  the  mixture  ratio  be-
tween  water  and  disinfectant.  It  is  located  at  the 
back  of  the  console.  It  can  only  be  calibrated  during 
a  disinfection  cycle  while  the  disinfectant  solution 
is  flowing.  The  table  next  to  it  should  help  you  to 
find  the  right  value.  Always  follow  the  instructions 
of  the  disinfectant  manufacturer.  The  Flowmeter  must 
be calibrated after each connection of water tubes and 
after changes in water pressure and after changing the 
disinfectant’s composition. During use of the tub, pay 
attention not to accidently alter its setting. 
 

Manually controlled disinfection procedure (model with 
disinfection valve) 

The  bathtub  may  only  be  disinfected  when  the  tub  is 
empty.  To  perform  the  disinfection  procedure  1)  take 
the 

red

  hand  shower  from  its  holder  and  2)  open  the 

disinfection

  valve  (in  counter  clockwise  direction). 

3)  Press  and  hold  the  trigger  directly  on  the  hand 
shower  and  spray  all  internal  tub  surfaces  with  the 
disinfectant solution. 
 

Caution: 

Before every disinfection procedure, ensure that there 
is enough disinfectant available in its container. 

 

 

 

Protect your skin and eyes from the disinfectant solu-
tion.  Follow  the  instructions  of  the  disinfectant 
manufacturer. 

 
4)  Close  the  disinfectant  valve  (in  clockwise  direc-
tion).  5)  Afterwards,  release  the  pressure  from  the 
hose  by  briefly  pressing  the  trigger  once  again  and 
6) put the hand shower back into its holder. 

After  the  necessary  contact  time  (according  to  the 
manufacturer’s  specifications)  the  bathtub  must  be 
rinsed  thoroughly.  To  do  that  7)  take  the 

white

  hand 

shower  from  its  holder,  8)  open  the 

shower

  valve  (in 

counter clockwise direction) and 9) press and hold the 
trigger  directly  on  the  hand  shower.  Rinse  all  inter-
nal  tub  surfaces  thoroughly.  Ensure  that  no  disinfec-
tant solution residues are left behind inside the tub. 
10)  Close  the  shower  valve  (in  clockwise  direction). 
11)  Afterwards,  release  the  pressure  from  the  hose  by 
briefly  pressing  the  trigger  once  again  and  12)  put 
the hand shower back into its holder. 

Содержание GK Magic 100

Страница 1: ......

Страница 2: ...and translation of both the whole document as well as of any parts thereof Congratulations on your decision to purchase a GK Bathtub You have acquired an Austrian quality prod uct which will considera...

Страница 3: ...ontrol panel 8 Shower 9 Shampoo Bath additive 9 Underwater light 9 Hydromassage 10 Sound Vision 10 Disinfection 11 12 Flowmeter 11 Manually controlled disinfection procedure model with disinfection va...

Страница 4: ...ce etc Never leave the patient unattended inside the bathtub Ensure that the equipment is thoroughly cleaned after every use When using disinfection protect your skin and eyes Heed all instructions on...

Страница 5: ...ure 5 Button Hydromassage lights up if configured 6 Button Down with printed arrow 7 Button Bath additive lights up if configured 8 Button Disinfection lights up if configured and needed Control panel...

Страница 6: ...Adjustable foot 8 Foot switch 9 Control panel 10 Control panel for Sound Vision 11 Tub fill valve 12 Thermostat mixing valve 13 Fill in spout 14 Shampoo Bath additive dispenser 15 Disinfection valve...

Страница 7: ...6 Components Hydromassage jets blue Box for care and disinfection liquids USB port Flowmeter with con centration table...

Страница 8: ...y means of the foot switch It works with air pressure and is connected to the electronics through thin air filled plastic tubes The higher air pressure when stepping on the switch activates the corres...

Страница 9: ...tion Always check the water temperature with your bare hand also when using the shower before bringing the water in contact with the patient Filling the tub Open the TUB FILL VALVE in counter clockwis...

Страница 10: ...a red marking on its suc tion end the tube for the bath additive a blue mark ing at the alike spot If there is no liquid coming out of the system make sure that both boxes are full After installing a...

Страница 11: ...2 mp3 wma is inserted into the USB port Now pressing the button PLAY PAUSE will play back all the recognized music files sorted by name Pressing the buttons NEXT and BACK switches to the next or previ...

Страница 12: ...irectly on the hand shower and spray all internal tub surfaces with the disinfectant solution Caution Before every disinfection procedure ensure that there is enough disinfectant available in its cont...

Страница 13: ...y all internal tub surfaces press the disinfection button again twice Thereby you are restarting the disinfection procedure and have an other 40 seconds available 5 After the 40 seconds are up and spr...

Страница 14: ...it to a comfortable working height Once the desired filling level has been reached stop if necessary the inflowing water To add bath additive to the water if installed press the button BATH ADDITIVE f...

Страница 15: ...p We especially recommend our special plastics intensive cleanser details on request For these parts of the tub only use very fine textured chamois leather or a microfiber cloth For the best care for...

Страница 16: ...ed cover of the thermostatic mixing valve Also the cartridge could be defective and might have to be changed The display reads F0 or F1 The temperature sensor is defective and must be changed Very lit...

Страница 17: ...d of the disinfection shower hose as well as of any externally visible connections Once a month check the bathtub for any cracks or other dam ages Caution Should you notice any damage to the bathtub e...

Страница 18: ...0 C Max relative humidity 80 not condensed at 20 C Atmospheric pressure range 500 hPa to 1060 hPa Transport and Storage Ambient temperature range 20 C to 60 C Relative humidity range 80 not condensed...

Страница 19: ...class I connection of protective earth conductor necessary Degree of protection IPX4 Protection type Type B Lifting weight 400 kg Lifting height 400 mm Maximum length 2240 mm Maximum width 930 mm Max...

Страница 20: ...Tel 43 0 7229 64450 Fax 43 0 7229 70664 service gk medizinmechanik at www gk medizinmechanik at Typing spelling or transmission errors excepted for all contents All product related information is not...

Отзывы: