Enter the programming and calibration menus by pressing the "CAL SWITCH" and entering the correct
password. (
,
and
) and press
key to enter
P1 Lin
Liner calibrate.
After you press
key, display will show
Pin
again, press
,
,
key and
key to enter,
display will show
LoAd 0
, remove all weight from the pan and wait until display show next step. Then
display will show
LoAd I
, put the weight follow the display, wait for next step.
Follow the display to load calibration weight on the pan.
Until display show
LoAd F
, load full capacity weight on the pan.
Note: calibration weight regulation: 0<load 1<load 2<load 3<load F(full capacity)
Load1, Load2, Load3 need be integer weight.
In TECH always press the push button and will be show the pin.
Version 1.16:
•
Press
key to enter calibration, display will be shown
PIN
•
Press
,
and
keys, display will be show
LOAD 0
(
*Note: Ensure the pan is empty
)
•
After stable and zero indicator on, display will be show
LOAD 1.
•
Place the first calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 2.
•
Add again the second calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 3.
•
Add again the third calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 4.
•
Add again the fourth calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 5.
•
Add again the fifth calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 6.
•
Add again the sixth calibration mass weight on the pan. (
*Note: Calibration mass weight
should be full capacity of the scale
)
•
After stable and zero indicator on, display will be show
LOAD 0.
Remove the calibration mass
weight from the pan. (
*Note: Ensure the pan is empty
)
•
After stable and zero indicator on, display will be show
LOAD 6.
•
Place the calibration mass weight on the pan (*
Note: Calibration mass weight should be full
capacity of the scale
)
•
After stable and zero indicator on, display will be show
LOAD 5.
•
Remove the sixth calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 4.
•
Remove the fifth calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 3.
•
Remove the fourth calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 2.
•
Remove the third calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 1.
•
Remove the second calibration mass weight on the pan.
•
After stable and zero indicator on, display will be show
LOAD 0.
•
Remove the first calibration mass weight on the pan.
(*Note: Ensure the pan is empty)
•
After stable and zero indicator on, display will be started self checking. After checking display will
come to normal weighing mode.
Note: During the linear calibration steps LOAD 1 to LOAD 6, mass weights want to add on the
pan. Don't remove all mass weights for the next step.
During the linear calibration steps LOAD 6 to LOAD 1, mass weights want to reduce from the
pan. Don't remove all mass weights for the next step.