Intelligente Datenlogger
- 2 -
© G.i.N. GmbH, 2020
SSD with exFAT format in a GL53xx
Please make sure that the system update is part of the COD file. For this, do
not
activate the parameter
“Do
NOT add system updates to code (-N)”
.
Figure 2
7.
Reconfigure the GL53xx with the new COD.
After a successful update, the GL53xx supports both file systems exFAT and FAT32.
8.
Remove the SSD from the GL53xx and connect it to the USB-eSATA adapter or insert it in the readout station.
9.
Launch the program MLcardReader (part of the MLtools/MLcenter package).
10.
Select the SSD as drive and press the “Format” button.
Please make sure to select the correct SSD when formatting, otherwise you may damage your system.
Figure 3