• Press the “Mode” button a second time “F-2” shown to set the alarm temper-

ature limit. Press the + or - buttons to change the value. 

  The mode is applied to Body Temp

•  Press the “Mode” button a third time “F-3” shown to set the °C/°F temperature 

units. The temperature units °C or °F will be shown. Press the + or - buttons to 

change the unit.

•  Press the “Mode” button a forth time “F-4” shown to set the backlit status.  

Press the + or - buttons to switch from ON to OFF.

•  Press the “Mode” button a fifth time “F-5” shown to set the laser status.  

Press the + or - buttons to switch from ON to OFF.

  Notice: Only when the laser head installed. 

Body Temperature Measurements

Please note that the forehead/temples must be free from sweat and cosmetics 

and that taking vasoconstrictive medication and skin irritations can distort the 

result when measuring temperature on the forehead. 
•  Press the trigger to switch the meter on with one beeper. The device performs 

a brief self test: all elements of the display are shown for approx. 1 second. 

Following a successful self test with backlit (Red-Orange-Green).

• Press the “B/O” button for “Body temp” for accurate measurements in the 

range of 32.0 to 42.9°C (89.6 to 109.2°F).

• Hold the meter by its handle and point it toward the surface to be measured. 

Measurement distance should be <2.5cm (<1 in).

• Press the trigger to take a temperature reading. The temperature reading ap-

pears in the display.

  If the F-4 MODE of backlit & “ 

 ” button voice ON & F-3 MODE set alarm 

temperature 38.0°C:

ENGLISH

Mode Function

“+”

“-”

Default

Remark

F-1

Setting Deviation 
Value

Increase

0,1°C

Decrease

0,1°C

0°C

Body mode available

Effective Range: ±2°C 

F-2

Setting Alarm Value

Increase

0,1°C

Decrease

0,1°C

38.0°C

Body mode available 

Effective Range: 

37.0-42.5°C

F-3

Setting Measurement

Unit

°F

°C

°C

F-4

Backlit ON/OFF

On

Off

On

F-5

Laser setting

On

Off

Off

Only when the laser 

head installed

Save

Save and automatic shut off

Содержание YI-400

Страница 1: ...N The operators must carefully read and completely understand the present manual before using the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produit ATEN...

Страница 2: ...and use a damp swab to wipe the lens Do not use any solvents to clean the lens Measurement Notes If the meter has been stored in a cold or hot environment allow it at least 30 minutes to acclimate to...

Страница 3: ...rature Mode settings With the meter ON press Mode button for three seconds until to F 1 shown to enter the long term calibration drift correction mode On entering the mode the previous temperature cor...

Страница 4: ...d Press the trigger to switch the meter on with one beeper The device performs a brief self test all elements of the display are shown for approx 1 second Following a successful self test with backlit...

Страница 5: ...re reading The temperature reading ap pears in the display Release the trigger and the reading will hold for approximately 15 seconds after which thermometer will automatically shut off Note Before sh...

Страница 6: ...specific recycling point for electric and electronic equipment For further information on recycling points contact the local authorities the local recycling center or the shop where the product was pu...

Страница 7: ...ctral response 8 to 14 um wavelength Display Backlit LCD display with function indicators Response time 500ms Over range indication Hi or Lo Operating Temperature Humidity 10 C to 40 C 50 F to 104 F 8...

Страница 8: ...ents ES Precauci n lea las instrucciones advertencias cuidadosamente IT Dispositivo medico conforme alla Direttiva 93 42 CEE GB Medical Device complies with Directive 93 42 EEC FR Dispositif m dical c...

Отзывы: