WARNING:

 For proper use, place the aspirator on a flat, stable surface in order to have the full 

volume of use of the jar and better efficiency of the overflow device. The vacuum jar, during use, must 

be used in vertical mode, to prevent the action of the backflow valve. If this protection is triggered, turn 

the device off and disconnect the pipe connected to the vacuum jar (indicated with the word VACUUM) 

on its cover.

WARNING:

 Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted 

into the “VACUUM”.

A wrong connection causes immediate destruction in case of contact with sucked liquids. 

FLOW DIRECTION

Suction Pump Inlet 

Jar Air Tube

Connect  the  short  silicon  tube,  with  filter

antibacterial, to the suction connector . The

other tube, with one end connected to the

filter must be connected with the other end

to jar's lid connector where has been fixed

the red float. When the 90% of the volume

of the jar is reached there is the activation

of  the  security  float  (the  float  close  the

aspiration connector on the jar) to avoid

liquid penetration inside the device. The

device must be used on a plan of horizontal

operation.

Connect the long silicon tube to the other

jar's lid connector

Connect the other end of the long silicon

tube to the probe plastic connector then

connect the suction probe to it.

Connect the power cord to the device then

connect the plug to the electrical mains

supply.

Push switch on position I to start suction

Once finished push switch on 0 position and unplug.

Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and

an rapid reaching of the functionally vacuum) then rescrew the lid on the jar correctly.

To extract the accessories and start with cleaning.

WARNING:

 The power supply cable plug is the element of separation from the electrical mains system: 

even if the units equipped with a special on / off switch button, the power supply plug must be kept ac-

cessible once the device is in use so as to allow a further method of disconnection from the mains supply 

system.

Содержание TOBI HOSPITAL

Страница 1: ...ASPIRATEUR TOBI HOSPITAL CHIRURGISCHER ABSAUGER TOBI HOSPITAL ASPIRADOR QUIR RGICO TOBI HOSPITAL 28202 28232 28233 28234 M28202 M Rev 7 01 21 Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gima...

Страница 2: ...unstable surfaces which may cause it to acci dentally fall and lead to a malfunction and or breakage Should there be signs of damage to the plastic parts which may expose inner parts of the energised...

Страница 3: ...ive and its normatives Under certain failure conditions the temperature of the casing may become hot and there may be a risk of burns if you touch those parts In any case the temperatures do not excee...

Страница 4: ...anteed up to 30 cycles of cleaning and sterilization Beyond this limit the physical chemical characteristics of the plastic material may show signs of decay Therefore we recommend that you to change i...

Страница 5: ...nd cooling at environment temperature of the parts make sure that these are not damaged The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121 C 1 bar relative pressure...

Страница 6: ...After sales Assistance Service If the overfill security system it s activated don t proceede with the liquid aspiration If the overfill security system doesn t work there are two cases 1 case If the...

Страница 7: ...e jar is reached there is the activation of the security float the float close the aspiration connector on the jar to avoid liquid penetration inside the device The device must be used on a plan of ho...

Страница 8: ...cover bends towards the inside of the glass the system is not leaking Start the aspiration and periodically check the filling level of the container The overflow valve will cause the interruption of a...

Страница 9: ...it s used in such an environment Immunity Test Level indicated by the EN 60601 1 2 Compliance Level Electromagnetic environments guidance Electrostatic discharge ESD EN 61000 4 2 8kV on contact 15kV i...

Страница 10: ...fixed RF transmitters as determined by an electromagnetic site study of the site should be less that the com pliance level in each frequency range Interference may occur in the vicinity of equipment...

Страница 11: ...3 10 3 8 3 8 7 3 100 12 12 23 For transmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to t...

Страница 12: ...specifications may change without notice Disposal The product must not be disposed of along with other domestic waste The users must dis pose of this equipment by bringing it to a specific recycling p...

Отзывы: