3. SpO2 Probe Connection
Connect the SpO2 probe to the right panel’s connector labeled “SpO2”. After
starting the monitor insert one finger into the probe (index finger, middle
finger or ring finger with proper nail length) according to demonstration, as
shown in the following figure.
Figure 5
Instructions of Operation
1. The finger should be put properly and correctly.
2. Do not shake the finger and keep at ease during using.
3. Do not put wet finger directly into sensor.
4. Avoid placing the sensor on the same limb which is wrapped with a cuff
for blood pressure measurement or during venous infusion.
5. Do not let anything block the emitting light from the sensor.
6. Vigorous exercise and electrosurgical device interference may affect the
measuring accuracy.
7. Using enamel or other makeup on the nail may affect the measuring
accuracy.