Operative Manual
MO082-IT
22/04/13
Rev.0
Pag. 11 di 12
9.
EMC declaration
The Product has been tested in accordance to EN60601-1-2 to ensure proper electromagnetic compatibility. Portable and
mobile RF-communications equipment can affect the Product. Other products used in the vicinity of Product should also
comply with this standard.
The Product is intended for use in the electromagnetic environment specified below. The customer or the user of the Product
should assure that these are used in such an environment.
Emissions test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The Product uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Product is suitable for use in all establishments other
than domestic, and may be used in domestic
establishments and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following
warning is heeded:
Warning:
This equipment/system is intended for use by
healthcare professionals only.
This equipment/system may cause radio interference or
may disrupt the operation of nearby equipment. It may be
necessary to take mitigation measures, such as re-orienting
or relocating the Product or shielding the location
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Compliance
Immunity test
IEC 60601-1-2
test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity should
be at least 30%
Electrical fast transient /
burst IEC 61000-4-4
+/- 2 kV
for power supply unit
+/- 1 kV
for input/output lines
+/- 2 kV
for power supply lines
+/- 1 kV
for input/output lines
Mains power quality should be that
of a typical commercial or residential
environment.
Surge
IEC 61000-4-5
+/- 1 kV
differential mode
+/- 2 kV
common mode
+/- 1 kV
differential mode
+/- 2 kV
common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
For 0,5 cycle
40% of U
T
(60% dip in U
T
)
For 5 cycles
70% of U
T
(30% dip in U
T
)
For 25 cycles
<5% U
T
(>95% dip in U
T
)
For 5 sec
<5% U
T
(>95% dip in U
T
)
For 0,5 cycle
40% of U
T
(60% dip in U
T
)
For 5 cycles
70% of U
T
(30% dip in U
T
)
For 25 cycles
<5% U
T
(>95% dip in U
T
)
For 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment.
If the user of the Product requires
continued operation during power
mains
interruptions,
it
is
recommended that the Product be
powered from an uninterruptible
power supply or battery.
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical
location
in
a
typical
commercial or hospital environment.
NOTE U
T
is the a.c. main voltage prior to application of the test level.