BATTERY REPLACEMENT
1. Replace battery when appears in the lower right corner of LCD display.
2. Pull battery cover off as shown in Figure 3.
3. Gently pull out plastic circuit board with battery chamber approximately 1
cm (slightly less than1/2
4. Use pointed object such as a pen to remove old battery. Discard battery
lawfully. Replace with new 1.5V DC button type LR41 or SR41,UCC392,
or equivalent. Be sure battery is installed with “+” polarity facing up. (See
Figure 5)
5. Slide battery chamber back into place and attach cover.
CALIBRATION
The thermometer is initially calibrated at the time of manufacture. If the ther-
mometer is used according to the use instruction, periodic readjustment is
not required. However, we recommend checking calibration every two years
or whenever clinical accuracy of the thermometer is in question. Turn on the
thermometer and insert into the water bath and then check the laboratory
accuracy of thermometer. Please send the complete device to the dealers
or manufacturer.
The above recommendations do not supersede the legal requirements. The
user must always comply with legal requirements for the control of the meas-
urement, functionality, and accuracy of the device which are required by the
scope of relevant laws, directives or ordinances where the device is used.
CLEANING AND DISINFECTION
Wipe the thermometer with a soft clean cloth.
For stubborn stains, wipe the thermometer with a cloth that has been damp-
ened with water or a neutral detergent solution and then wring thoroughly.
Finish by wiping with a soft dry cloth.
For disinfection, 75% Ethanol or Isopropyl alcohol can be used.
Observe the following to prevent damage to the thermometer.
Do not use benzene, thinner, gasoline or other strong solvents to clean the
thermometer.
Do not attempt to disinfect the sensing section (tip) of the thermometer by
immersing in alcohol or in hot water (water over 50℃ (122℉).
Do not use ultrasonic washing to clean the thermometer.
Disposal: The product must not be disposed of along with other do-
mestic waste. The users must dispose of this equipment by bringing
it to a specific recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
The system is not
functioning properly.
Unload the battery, wait for
1 minute and repower it. If the
message reappears, contact the
retailer for service.
Dead battery: Battery
icon is flashing, can’t be
measurable.
Replace the battery.
Type BF applied part
Follow instructions for use
Manufacturer
Date of manufacture
Keep in a cool,
dry place
Product code
Keep away from
sunlight
Lot number
Authorized represent-
ative in the European
community
Caution: read instructions
(warnings) carefully
WEEE disposal
Medical Device complies
with Directive 93/42/EEC
Temperature limit
IP27
Covering Protection rate
Humidity limit
Direct current
Atmospheric pressure
limit