PROFESSIONAL MEDICAL PRODUCTS
PLEASE READ CAREFULLY BEFORE USING
This digital thermometer provides a quick and highly accurate reading of an
individual’s body temperature. The digital thermometer is intended to meas-
ure the human body’s temperature in regular mode orally, rectally or under
the arm, and the device is reusable for clinical or home use on people of all
ages. To better understand its functions and to provide years of dependable
results, please read all instructions first.
This appliance conforms to the following standards:
EN 12470-3 Clinical thermometers —Part 3: Performance of compact elec-
trical thermometers (non-predictive and predictive) with maximum device,
ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular require-
ments for basic safety and essential performance of clinical thermometers
for body temperature measurement, EN 60601-1-11 Medical electrical
equipment —Part 1-11: General requirements for basic safety and essen-
tial performance – Collateral Standard: Requirements for medical electrical
equipment and medical electrical systems used in the home healthcare en-
vironment and complies with the requirements of EN 60601-1-2(EMC), IEC/
EN60601-1(Safety) standards. And the manufacturer is ISO 13485 certified.
Warning:
• Read instructions thoroughly before using digital thermometer.
• Choking Hazard: Thermometer cap and battery may be fatal if swallowed.
Do not allow children to use this device without parental supervision.
• Do not use thermometer in ear. Designed use is for oral, rectal, and armpit
(axilla) readings only.
• Do not place thermometer battery near extreme heat as it may explode.
• Remove battery from the device when not in operation for a long time.
Note: Use of the probe cover may result in a 0.1℃(0.2℉) discrepancy
from actual temperature.
• The use of temperature readings for self-diagnosis is dangerous. Consult
your doctor for the interpretation of results. Self-diagnosis may lead to the
worsening of existing disease conditions.
• Do not attempt measurements when the thermometer is wet as inaccurate
readings may result.
• Do not bite the thermometer. Doing so may lead to breakage and/or injury.
• Do not attempt to disassemble or repair the thermometer. Doing so may
result in inaccurate readings.
• After each use, disinfect the thermometer especially in case the device is
used by more than one person.
• Do not force the thermometer into the rectum. Stop insertion and abort the
measurement when pain is present. Failure to do so may lead to injury.
• Do not use thermometer orally after being used rectally.
• For children who are two years old or younger, please do not use the de-
vices orally.
• If the unit has been stored at temperatures over 5℃~40℃(41℉~104℉),
leave it in 5℃~40℃ (41℉~104℉) ambient temperature for about 15 min-
utes before using it.
PRECAUTION
The performance of the device may be degraded should one or more of the
following occur:
- Operation outside the manufacturer’s stated temperature and humidity
range.
- Storage outside the manufacturer’s stated temperature and humidity
range.
- Mechanical shock (for example, drop test) or degraded sensor.
- Patient temperature is below ambient temperature.
* Portable and mobile RF communications can affect the device. The device
needs special pre-cautions regarding EMC according to the EMC informa-
tion provided in the accompany documents.
SPECIFICATIONS
DIRECTIONS
1. Press the On/Off Button next to LCD display. A tone will sound as the
screen shows
, followed by last recorded temperature. After show-
ing the self-test temperature, the thermometer is now in the testing mode.
2. Position thermometer in desired location (mouth, rectum, or armpit.)
a) Oral Use: Place thermometer under tongue as
indicated by “√” position shown in Figure 2. Close
your mouth and breathe evenly through the nose to
prevent the measurement from being influenced by
inhaled/exhaled air. Normal temperature between
35.7℃ and 37.3℃ (96.3℉ and 99.1℉). Figure 2
b) Rectal Use: Lubricate silver probe tip with petrole-
um jelly for easy insertion. Gently insert sensor approximately 1cm (less
than 1/2″) into rectum. Normal temperature between 36.2 and 37.7℃
(97.2℉ and 99.9℉).
c) Armpit Use: Wipe armpit dry. Place probe in armpit and keep arm
pressed firmly at side. From a medical viewpoint, this method will al-
ways provide inaccurate readings, and should not be used if precise
measurements are required. Normal temperature between 35.2℃ and
36.7℃(95.4℉ and 98.1℉).
3. The degree sign flashes throughout the testing process. When flashing
stops an alarm will beep for approximately 10 seconds. The measured
reading will appear on the LCD simultaneously. The minimum measure-
ment time until the signaling tone (beep) must be maintained without ex-
ception. The measurement continues even after the buzzer notification.
So that in order to achieve better body temperature measurement result,
recommend to keep the probe in mouth and rectum about 2 minutes, or
in armpit about 5 minutes regardless of the beep sound and at least 30
seconds measurement interval should be maintained.
*Note: Normally the buzzes are“ Bi-Bi-Bi- Bi-”;Alarm beeps more rap-
idly when temperature reaches 37.8 (100℉) or higher ,and the
buzzes are“ Bi-Bi-Bi----- Bi-Bi-Bi------- Bi-Bi-Bi”
4. To prolong battery life, press the On/Off Button to turn unit off after testing
is complete. If no action is taken, the unit will automatically shut off after
around10 minutes.
TROUBLESHOOTING
Probe
ON/Off Button
LCD
Battery Cap
Error
message
Problem
Solution
Temperature taken is
lower than 32.0°C
(90.0°F)
Turn off, wait one minute and
take a new temperature via
close contact and sufficient rest.
Temperature taken is
higher than 42.9°C
(109.9°F)
Turn off, wait one minute and
take a new temperature via
close contact and sufficient rest.
ENGLISH
Type:
Digital Thermometer (Not Predictive)
Measure Range:
32.0℃ - 42.9℃° (90.0℉-109.9℉ )
(℃/℉ chosen by manufacturer)
Accuracy:
±0.1℃ (±0.2℉) during 35.5℃~42.0℃
(95.9℉~107.6℉) at 18℃~28℃
(64.4℉~82.4℉) ambient operating range
±0.2℃ (±0.4℉) for other measuring and
ambient operating range
Operating mode:
Direct Mode
Display:
Liquid crystal display, 3 1/2 digits
Memory:
For storing the last measured value
Battery:
One 1.5 V DC. button battery
(size LR41or SR41, UCC 392)
Battery life:
Approx. 200hours of continuous operation
or 1 year with 3 measurements per day
Dimension:
13.9cm×2.3cm×1.3cm (L×W×H)
Weight:
Approx. 13 grams including battery
Expected service life:
Three years
Ambient operating range: Temperature: 5℃~40℃(41℉~104℉)
Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa
Storage and transporta-
tion condition:
Temperature: -20℃~55℃(-4℉~131℉)
Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa
Ingress Protection Rating: IP 27
Classification:
Type BF